Service manual
CHAPTER 1MAINTENANCE TESTS
34 9650-0801-01 Rev. A
NIBP Test
Tools Needed NIBP simulator, single or dual lumen hose, adult, NIBP cuff, rigid cylinder
Test Setup Connect the Propaq MD to the NIBP simulator using the single or dual lumen hose. Wrap the 
NIBP cuff around a rigid cylinder, such as a glass jar.
Note: Refer to Appendix B for information on simulators and settings that you may use to 
assess the performance of the NIBP module.
Do This... Observe This... Pass/Fail
1. On the Propaq MD, ensure the patient 
mode is set to Adult.
2. Set the simulator to the 80/50 simulator 
preset.
3.  Take a manual reading. Confirm that the NIBP measurement 
reported is within +/- 10 mmHg of 
the expected simulator performance.
oo
4. Set the simulator to the 120/80 simulator 
preset.
5. Take a manual reading. Confirm that the NIBP measurement 
reported is within +/- 10 mmHg of 
the expected simulator performance.
oo
6. Set the simulator to the 200/150 simulator 
preset.
7. Take a manual reading. Confirm that the NIBP measurement 
reported is within +/- 10 mmHg of 
the expected simulator performance.
oo










