Service manual

9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 5
Chapter 1
Maintenance Tests
Propaq MD Overview
This chapter includes step-by-step procedures for the following types of maintenance tests:
Calibration and Calibration Checks
NIBP Valve, Pump, and Tubing Integrity Tests
Operational Tests
Functional Verification Tests
At a minimum, a trained biomedical technician must perform a CO
2
calibration and an NIBP
calibration check once per year. A CO
2
calibration check may be performed at any time to
verify the CO
2
calibration.
The Troubleshooting chapter in this manual may recommend performing one of more of the
NIBP valve, pump, and tubing integrity tests to help identify and correct potential issues with
the NIBP valve, pump, and/or tubing.
It is necessary to perform Functional Verification Tests after a module is repaired to ensure that
the functions of the Propaq MD unit work properly.
Because the Propaq MD units must be maintained ready for immediate use, it is important for
users to conduct the Operators Shift Checklist procedure at the beginning of every shift. This
procedure can be completed in a few minutes and requires no additional test equipment. (See
the ZOLL Propaq MD Operator’s Guide for the Operators Shift Checklist.) It is highly
recommended that the Operational Tests be performed at regular intervals to augment the
automatic testing the Propaq MD does to ensure readiness.