Specifications
11
Pacemaker Analyzer
Key features
• Defibrillator analyzer
• Monophasic and biphasic dc 
energy measurement
• Energy and cardioversion 
measurement
• Peak voltage, peak current, and 
overshoot measurement
• 2-line x 24-character display
• Bidirectional RS-232 port for 
computer control
• Storage and playback of output 
waveform so results can be 
viewed in greater detail
• Optional Ansur test automation 
software to standardize testing 
procedures, capture waveforms 
and test results, and print and 
document test results
Fluke Biomedical’s premier 
SigmaPace 1000 analyzes both 
transvenous and transcutaneous 
external pacemakers and comes 
loaded with features to save time 
and money. This powerful hand-
held tool conducts the full suite of 
tests specified by major pacemaker 
manufacturers in less than half the 
time it would take using originally 
prescribed testing methods.
Output data is displayed on three 
selectable screens for 
easy viewing, including an AV 
delay time readout providing a 
performance snapshot for both 
pacer channels.
ECG disposable snap electrode 
adapters
3.2 mm and 4 mm
Modes of operation Manual, remote
Transcutaneous pacer tests Pulse output test 
Current: 4 mA to 250 mA 
Rate: 5 PPM to 300 PPM
Width: 1 ms to 99.9 ms
Energy: 1 µJ to 1.99J
Demand model test
Async mode test
Amplitude sensitivity test
Noise immunity test
Paced refractory period test
Sensed refractory period test
Internal test loads: 31 selections: 50 Ω to 1550 Ω
SigmaPace
™
 1000
External Pacemaker Analyzer
Specications
With capability for long-term trend testing, the SigmaPace 1000 can 
interrogate a pacer for up to 11 days, capturing data pulse by pulse to 
detect intermittent and hard-to-find problems.
For maximum efficiency, the SigmaPace 1000 doubles as a training tool. 
Interactive ECG simulation lets users test patient monitoring equipment as 
well as teach nurses how to operate the pacemaker. 