Specifications
12 Pacemaker Analyzer
Optional accessories
9513-0202 Electrode adapters 
(including the brand/model-specific 
interface connector and a pair of 4 mm 
“safety-type banana plugs)
3010-0605 Aligent (HP) CodeMaster 
Series
3010-0606 GE Marquette Medical
3010-0607 Medical Data Electronics 
(MDE); Medical Research Laboratories 
(MRL)
3010-0604 Medtronic Physio-Control 
Quick Combo
3010-0603 Medtronic Physio-Control 
Quick Pace
3010-0639 Philips/Agilent 
Codemaster Series
3010-0608 Zoll Medical NTP Series
3010-0609 Zoll Medical PD Series and 
M Series
3010-0441 Interface Cable (RS-232; 
female DB9 to female DB25; medTester 
to SigmaPace™ 1000/PC/Index 2XL/
IDA 4 Plus; Impulse 4000 to PC)
3010-0654 Detachable cord set—Japan 
(IEC 320 C6 type 3-pin inlet)
3010-0656 Detachable cord set— 
Schuko—Euro (IEC 320 C6 type 3-pin 
inlet)
3010-0655 Detachable cord set—UKI 
(IEC 320 C6 type 3-pin inlet)
3010-0658 Detachable cord set—USA 
(IEC 320 C6 type 3-pin inlet)
3010-0657 Detachable cord set— 
Australia (IEC 320 C6 type 3-pin inlet)
Ordering information
SIGMAP1K-USA120V United States, 
120 V
SIGMAP1K-JPN100V Japan, 100 V
SIGMAP1K-SHK250V Schuko, 250 V
SIGMAP1K-UK250V United Kingdom, 
250 V
Included accessories
9508-0295 Users Manual
9530-0069FG Soft-sided Vinyl 
Carrying Case
3010-0611 Transvenous Pacer Test 
Leads (2 sets, red)
3010-0610 Transvenous Pacer Test 
Leads (2 sets, black)
3010-0602FG SigmaPace 9 V dc Load 
Test Cable
3010-0585FG Serial PC Interface Cable
POWER SUPPLY Universal-input 
Battery Charger
LINE CORD Power cord set USA  
120 V ac
Transvenous pacer tests Pulse output test
Display output test: (3) single (A or V) and dual (A+V)
Current: 0.05 mA to 30 mA
Rate: 10 PPM to 999 PPM
Width: 0.02 ms to 9.99 ms
Voltage: 0.05 V peak to 30 V peak
Energy: 1 nJ to 999 µJ
Demand model test
Async mode test
Amplitude sensitivity test
Noise immunity test
Paced refractory period test
Sensed refractory period test
AV delay time test
DC leakage current test
Measurement of dc offset on the pacemaker output
Test types: Static/continuous and dynamic/sync’d with output
Inputs/test loads: (3) 500 Ω
DC current range: 0.1 mA to 99.9 mA
Battery load current test
Load current drawn by the pacemaker: 99.9 mA max
Transvenous 
measurement algorithm
Default: Derived from device manufacturer(s)
Available internal test 
loads
Ventricular and atrial channels: 200 Ω, 500 Ω, and 1000 Ω
Default selection: 500 Ω (both A and V channels)
Long-term trend test Tests the fundamental stability of the pacer output
Total pulse count: 999,999 (max)
Elapsed time: 999:59:59 (max)
Maximum error count: 200 pulses
Test limits: Selectable rate and output percentages
Interactive pacer ECG 
simulation
Simulates dynamic patient ECG activity in response to pacer 
output: 
Interactive NSR heart rate: 0 BPM to 25 BPM
NSR PR interval: 0.05 s to 0.399 s
Transvenous threshold: 1 mA to 25 mA
Transcutaneous threshold: 10 mA to 250 mA
Serial port Type: RS-232
Baud rate: 2400, 9600, and 192000
Power Internal lithium ion battery pack (rechargeable)
Battery operation: 20 hours (minute)
Universal/external ac-to-dc power supply
Dimensions (WxDxH) 10.1 cm 20.3 cm x 5 cm (4 in x 8 in x 2 in)
Weight 0.9 kg (2 lb)
Specications
SigmaPace
™
 1000
External Pacemaker Analyzer
riešenia na presné meranie
http://www.elso.sk










