Specifications
15
Electrosurgery Analyzers
Key features
• Easy to use simple 
configuration
• Oscilloscope output, high-
frequency leakage, and return 
electrode contact quality moni-
tor tests 
• Instantaneous and signal- 
averaging measurement mode
• Ability to connect with Fluke 
Biomedical’s medTester 5000C 
for automated solution
• RS-232 port for computer 
control
• Battery powered
• 4-digit numeric LCD with back-
light and power-save mode
RF303
RS
 Electrosurgery 
Analyzer provides enough 
user-selectable test loads 
to do routine maintenance 
checks on most electrosur-
gery units on the market 
today. Compact and por-
table, the device is so 
simple to use that techni-
cians can become proficient 
with the RF303
RS
 within 
minutes.
The unit measures ESU 
output and high-frequency 
leakage, allows for verifi-
cation tests on the return electrode contact quality monitors, and has an 
oscilloscope output for waveform viewing. Instantaneous output or select-
able sample times provide extra versatility. The instantaneous mode is 
sufficient for most units, but if output readings are variable and require 
stabilizing, the signal-averaging mode allows users to manually select two 
additional, slower sampling times to produce an accurate average reading.
Modes of operation Line powered, battery powered, offline (battery maintenance 
charge)
Test parameters
Power (W), HF current (mA), test load (Ω)
Tests performed 
generator output
HF leakage (performs to IEC 601 2-2, 1289-2, ANSI/AAMI standards)
Type BF test 1: Earth-referenced monopolar output
Type BF test 2: Earth-referenced monopolar output
Type CF/bipolar: Isolated monopolar or bipolar output
Current measurement 
(leakage)
Range: 30 mA to 2500 mA rms
Resolution: 1 mA
Accuracy: ± 2.5 % of reading or ± 15 mA (whichever is greater)
Power measurement 
(output)
Range: 1 W to 400 W
Resolution: 0.1 W
Accuracy: ± 5 % of reading or ± 3 W (whichever is greater)
Bandwidth of rms 
converter circuit 
(1 % accuracy)
Flat response: 10 kHz to 10 MHz
-3 dB points: 1 kHz to 20 MHz
Frequency response
System response: -3 dB points, 1 kHz to 10 MHz at 300 Ω
RECQM test
50 Ω to 750 Ω, 50 Ω steps
Test load section Number of selections: 15
Range: 50 Ω to 750 Ω
Step size: 50 Ω
Accuracy (dc to 500 KHz): ± 4 % of selected value measured at 
calibration to ± 1 % (across the entire operating temperature range)
Duty cycle: 25 % @ 400 W (max 30 sec ON during any 2 minute period)
Resonance impedance variation: ± 0.5 dB max (< 10 MHz)
Auxiliary leakage test 
load
Fixed: 200 Ω
Accuracy: ± 4 %
Power rating: 225 W
Input capacitance
(nominal)
Active to dispersive: 30 pF
Active or dispersive to earth ground: 40 pF
Oscilloscope output Transformer coupled output, uncalibrated
Connector type: BNC
RF303
RS
Electrosurgery Analyzer
Specications