Specifications
19
Electrical Safety Analyzers
Key features
• Superior compliance with 
multiple standards: IEC60601-
1(partial), IEC62353, VDE 751, 
ANSI/AAMI ES1:1993, NFPA-99, 
AN/NZS 3551, IEC61010
• Three test loads
• Expanded leakage ranges 
through 10,000 µA
• Dual-lead resistance, leakage, 
and voltage tests
• AC only, dc only and true-rms 
leakage readings
• 100 % and 110 % mains volt-
age for mains on applied part 
(lead isolation) test
• DSP filter technology for 
improved accuracy in leakage 
measurements
• 20 A equipment current
• More applied parts selections
• ECG and performance 
waveforms
• Intuitive user interface
• Easy-to-use applied parts (ECG) 
connections
• Insulation posts on applied 
parts connections
• Five different insulation tests
• Varying insulation test voltage 
500 V dc and 250 V dc
• 2- or (optional) 4-wire ground 
wire resistance
• Large display with adjustable 
contrast
• Ergonomic design
• Optional Ansur plug-in 
software
• USB connection
• CE, C-TICK and CSA for USA 
and Canada
• RoHS compliance
• Optional Ansur test automation 
software to standardize testing 
procedures, capture waveforms, 
and print and document test 
result
The ESA620 Electrical 
Safety Analyzer repre-
sents the next generation 
in manual, portable elec-
trical safety testers. With 
selections of three test 
loads, protective earth 
test currents, and two 
insulation test voltages 
this versatile product 
can be used worldwide 
to enhance productivity 
and test to standards of 
choice.
New DSP technology offers better accuracy of leakage measurements 
throughout the ranges specified in the standards.
Equipped with ten safety-enhanced ECG posts, the ESA620 offers 
simulation of ECG and performance waveforms so both electrical safety 
and basic tests on patient monitors can be performed with a single con-
nection. When used with optional Ansur computer-based software plug-in, 
the ESA620 becomes automated. This allows for standardization of test 
procedures, capturing and storage of results, comparison to standard limits, 
and printing of reports thus enabling the sophisticated performance of the 
high-end electrical safety analyzers.
Voltage
Range (mains voltage) 120 V model: 90 V ac to 132 V ac rms
230 V model: 180 V ac to 264 V ac rms
Accuracy ± (2 % of reading + 1 V)
Range (accessible voltage) 0 V ac to 300 V ac rms
Accuracy ± (2 % of reading + 2 LSD)
Voltage tests Mains, Accessible, and Point to Point
Earth resistance
Modes Two terminal or four terminal
Test current > 200 mA ac or 10 A ac to 25 A ac
Ranges
0 Ω to 2 Ω
Accuracy
± (2 % of reading 0.015 Ω)
Equipment current
Mode AC rms
Range 0 A to 20 A
Accuracy ± 5 % of reading ± (2 counts or 0.2 A, whichever is greater)
Leakage current
Patient load selection 
(input impedance)
AAMI ES1-1993 Fig 1 
IEC 60601: Fig 15 
IEC 61010: Fig A-1
Crest factor ≤ 3 
Ranges 0 μA to 199.9 μA
200 μA to 1999 μA
2.0 μA to 10.0 mA
Frequency response DC to 1 kHz
1 kHz to 100 kHz
100 kHz to 1 MHz
Accuracy ± (1 % of reading + 1 μA or 1 LSD, whichever is greater)
± (2 % of reading + 1 μA or 1 LSD, whichever is greater)
± (5 % of reading + 1 μA or 1 LSD, whichever is greater)