Specifications
26
Optional accessories
97148 Soft-sided Vinyl Carrying Case
600/100FG Chassis Cable, coil cord, 
8-foot extended
600/101FG Chassis Cable, coil cord, 
16 foot extended
600/200FG External Leakage Cable, 
coil cord, 8 foot extended
600/201 External Leakage Cable, coil 
cord, 16 foot extended
1800001 North American 220 V 
adapter kit
Ordering information
ESA175-US 90-265V 175 Electrical 
Safety Analyzer
Included accessories
1751000 Operators Manual
600/100FG Chassis cable, coil cord,  
8 ft extended
Key features
• Portable
• IEC601-1 and AAMI test loads
• Self-switching 120 V and  
240 V operation
• Dual-lead testing
The 175 Electrical Safety 
Analyzer is ideal for performing 
quick electrical safety checks 
on electrical systems, medical 
devices, and physiological 
instrumentation.
Small enough to fit in a brief-
case, the analyzer works well 
as a bench-top instrument in 
the laboratory or a portable 
testing device in the mobile 
engineer’s toolbox.
With its uncomplicated 
design, the 175 is simple to 
use. A single master function switch, directly labeled with the test to be 
performed, leads the user through a complete measurement procedure. The 
unit features both IEC601-1 and AAMI test loads. Technicians simply flip a 
switch to perform leakage measurements to a particular standard.
The analyzer uses simple, yet sophisticated, electronics for true-rms 
measurement of current and voltage. The 175 also performs dual-lead 
leakage tests.
Operating mains 
voltage range
90 V ac to 240 V ac
Current-capacity for DUT 15 A for 10 min
Voltage measurement
Range 90 V to 240 V
Accuracy ± 3 % of reading ± 1 LSD
Load-current measurement
Range 1 A to 19.99 A
Accuracy ± 5 % of reading ± 1 LSD
Resistance measurement
Range
0.01 Ω to 19.99 Ω
Accuracy ± 1 % of reading ± 1 LSD
Resolution
0.01 Ω
Current source 10 mA dc
Leakage-current measurement
Range 0 µA to 1999 µA
Accuracy DC and 25 Hz to 1 KHz: ± 1 % of reading ± 1 µA; 1 KHz 
to 100 KHz: 
± 2.5 % of reading ± 1 µA; 100 KHz to 1 MHz: ± 5 % of 
reading ± 1 µA
Measurement type True-rms
Input impedance
1000 Ω per AAMI ES1-1993, IEC601-1
Environmental requirements 
Operating temperature 15 °C to 40 °C (59 °F to 104 °F)
Storage temperature -20 °C to 50 °C (-4 °F to 122 °F)
Relative humidity Max 80 % up to 31 °C (88 °F), decreasing linearly to 50 % at 
40 °C (104 °F)
General information
Display LCD 3.5 digit
Dimensions (LxWxH) 13.3 cm x 18.4 cm x 5.4 cm (5.25 in x 7.25 in x 2.1 in)
Weight Max weight: 1 kg (2.25 lb)
175
Electrical Safety Analyzer
Specications
Electrical Safety Analyzers