Specifications
30 Non-Invasive Blood Pressure Simulators
Key features
• Dynamic BP simulators for arm- 
and wrist-cuff monitors
• Physiological waveform
• Internal pump for high- and 
low-pressure release verifica-
tion, leak testing, and pressure 
sourcing
• Preset mode for simulation of 
most patient conditions
• Five-leads synchronized ECG 
simulation
• Arrhythmia simulations, 
including premature atrial 
contractions #1 and #2, atrial 
fibrillation, and PVCs
• Optional Ansur test automation 
software to standardize testing 
procedures, capture waveforms, 
and print and document test 
result
Optional accessories
ANSUR BP PUMP 2 Ansur BP Pump 2 
Plug-in
75034 Data Transfer Cable
5022010 Soft-sided Vinyl Carrying 
Case
2780512FG ECG Adapter Block (allows 
simulation of 5-lead ECG waveforms)
PRINTR/414-US120V Printer, Seiko 
DPU-414-30 B, 120 V power supply
PRINTR414-SHK220V Printer, Seiko 
DPU-414-30B, 200 V power supply
61096 Printer, 120 V power supply
61097 Printer, 220 V power supply
75034 Serial Cable, D9M-D9F
71072 Parallel Printer Cable, D25M-
C36M
5027-0203FG Neonatal/external cuff 
mandrel (truncated plastic cylinder 
diameters: 7.6, 10, and 14 cm)
98175FG Wrist cuff mandrel (adult)
5215-0269FG Adult Cuff Mandrel 
Spacer Block (three required)
5215-0268FG Adult Cuff Mandrel End 
Block (two required)
BPPM2M-2001 500 ML Rigid 
Aluminum Chamber for NIBP Testing
BPPM2M-2002 100 ML Rigid 
Aluminum Chamver for NIBP Testing
700PMP External Pressure Pump
Quick NIBP monitor testing bundles
PS420/DPM1B Bundle PS420/DPM1B 
Bundle Kit (includes PS420, DPM1B, 
all accessories, and a custom carrying 
case)
PS410/DPM1B Bundle PS410/DPM1B 
Bundle Kit (includes PS410, DPM1B, 
all accessories, and a custom carrying 
case)
Pressure generation/ 
measurement
Static-pressure range: 50 mmHg to 400 mmHg (53 kPa)
Difference between target pressure and actual pressure: -5 mmHg
Internal leak rate: < 2 mmHg per minute with minimum volume of 300 cc
Four respiratory artifacts 3 spontaneous breathing; controlled ventilation
Three adult wrist-cuff simu-
lations
Normal, Hyper, Hypo
Pressure source Specified pressure generated from 50 mmHg to 400 mmHg in selectable incre-
ments of 1 mmHg
Pressure gauge Static pressure measured from 0 mmHg to 400 mmHg at the pressure port
Pressure relief test Test for the NIBPM pressure relief valve (0 mmHg to 400 mmHg) with display of 
peak pressure
Neonate internal cuff simu-
lations
Internal neonate cuff; four standard neonate pressures
Neonate simulations Cuff #1:  Blood pressure: 35/15  Heart rate: 120 BPM  Pulse volume: 0.1 cc
Cuff #2:  Blood pressure: 60/30  Heart rate: 120 BPM  Pulse volume: 0.1 cc
Cuff #3:  Blood pressure: 80/50  Heart rate: 120 BPM  Pulse volume: 0.1 cc
Cuff #4:  Blood pressure: 100/70  Heart rate: 120 BPM  Pulse volume: 0.1 cc
Irregular pulse BP and ECG: Premature atrial contractions # 1, premature atrial contractions # 2, 
premature ventricular contractions, atrial fibrillation and PVCs
User-definable simulations User-definable systolic and diastolic values, along with heart rate and pulse volume 
Ranges:
Systolic pressure  Diastolic pressure  Heart rate  Pulse volume
range range
20 mmHG to  10 mmHG to  30 BPM to  0.1 cc to 2.4 cc in 
250 mmHG  200 mmHG  250 BPM  increments of 0.1 cc
Simulation parameters 
performance
Max pulse volume: 2.4 cc
Max heart rate: 200 BPM at 2.4 cc pulse volume; 250 BPM at 1.2 cc pulse volume
Internal neonatal cuff volume: 20 cc
Internal adult cuff volume (including NN volume): 310 cc
Heart rate setting accuracy: ± 1 BPM
Simulation units: kPa and mmHg (user selectable)
Pressure leak test The pressure port is pressurized from 0 mmHg to 400 mmHg and keeps track of the 
pressure loss over time. Peak pressure and present pressure are displayed at all 
times; leak rate is displayed when it is available.
Autosequences Nine autosequences are provided for four tests and up to five simulations
Electrical ECG (optional) Signals: RA, LA, RL, LL, V
Waveform: Lead II
Amplitude: 1 mV peak (± 10 %) NIBP peripheral pulse synchronized with ECG signal
Connections: Optional external ECG adapter, physiological synchronization with 
NIBP
The BP Pump 2 is a second-
generation non-invasive 
blood pressure (NIBP) moni-
tor analyzer that efficiently 
verifies oscillometric adult and 
neonatal NIBP. The BP Pump 
2’s unique feature set includes 
tests to accurately interrogate 
wrist-cuff monitors, internal 
cuff volumes, and optional 
5-lead synchronized ECG sim-
ulations for spot checks on the 
monitor. The simulated periph-
eral pulse is synchronized with 
this electrical ECG signal for 
testing NIBP monitors utilizing 
gated measurement for noise/
artifact rejection.
BP Pump 2
Non-Invasive Blood Pressure Simulator
Specications