Specifications
43
Patient Simulators
Blood pressure 
channels
Channels 
4; synchronized with normal sinus rhythm rates; tracks arrhythmia 
activity
Transducer
Exciter voltages: ac and dc compatible
Sensitivity: 5 μV/V/mmHg and 40 μV/V/mmHg
Calibrated Rate: 80 BPM
Available selections Static pressure
Dynamic pressure: Art (120/80), Radial Art (120/80), LV (120/0), RA/
CVP (15/10), RV (25/0), PA (25/10), PAW (10/2), and LA (14/4)
Swan-Ganz procedure: automated and manual control
Temperatures 0 °C, 24 °C, 37 °C, and 40 °C
Cardiac output 
Thermodilution 
method, (optional)
Faulty-injectate curve
Left-to-right shunt curve
C.O. for 0°: 2.5 l/min, 5 l/min, and 10 l/min
C.O. for 24°: 2.5 l/min, 5 l/min, and 10 l/min
Cal Pulse: 1.5° for 1 second
Fetal/Maternal ECG 
and IUP 
simulations 
(optional)
Fixed fetal heart rates: 60 BPM to 240 BPM
Dynamic fetal heart activity: Acceleration, early deceleration, late 
deceleration
Maternal heart rate: 80 BPM
Dynamic intrauterine pressure (IUP)
Waveform: positive bell-shaped pressure curve
Peak pressure: 90 mmHg, ± 4 mmHg (max)
Contraction interval: 2 minutes, 3 minutes, and 5 minutes (manual)
Duration: 90 seconds
Dimensions (WxDxH) 18.4 cm x 19 cm x 5 cm (7.3 in x 7.5 in x 2 in)
Weight 0.6 kg (1.4 lb)
MPS450
Patient Simulator
Specications
Optional accessories
9530-0072 Soft-Sided Carrying Case
75034 Serial Cable D9M-D9F
5180500 Cardiac-Output Adapter Box
HHC3 Handheld Controller
Cardiac output adapters
3010-0650FG GE Medical/Marquette 
Cardiac Output Cable (interface cable 
for GE Medical/Marquette plus moni-
tors, including in-line switch box to 
select injectate temperature)
3010-0285 HPT-2 Temperature adapt-
er (Hewlett Packard/Philips) (2 pin)
3010-0284 COA-1 Cardiac Output 
adapter (HPT-2 also required for 
cardiac output simulation on patient-
monitoring systems)
5183004FG Universal Injectate 
Temperature Adapter Pigtail 
(unterminated)
For a complete list of temperature and blood-
pressure cables, contact us.
Ordering information
MPS450 MPS450 Patient Simulator
MPS450-01 Australia
MPS450-02 Denmark
MPS450-03 India
MPS450-04 Israel
MPS450-05 Italy
MPS450-06 Schuko
MPS450-07 Switzerland
MPS450-08 United Kingdom
MPS450-09 Brazil
MPS450-CO MPS450-CO (base model 
plus cardiac-output simulation)
MPS450-CO-01 Australia
MPS450-CO-02 Denmark
MPS450-CO-03 India
MPS450-CO-04 Israel
MPS450-CO-05 Italy
MPS450-CO-06 Schuko
MPS450-CO-07 Switzerland
MPS450-CO-08 United Kingdom
MPS450-CO-09 Brazil
MPS450-FET MPS450-FET (base 
model plus direct fetal/maternal ECG 
simulations
MPS450-FET-01 Australia
MPS450-FET-02 Denmark
MPS450-FET-03 India
MPS450-FET-04 Israel
MPS450-FET-05 Italy
MPS450-FET-06 Schuko
MPS450-FET-07 Switzerland
MPS450-FET-08 United Kingdom
MPS450-FET-09 Brazil
MPS450-CO/FET MPS450-CO/FET (base 
model plus cardiac-output and direct 
fetal/maternal ECG simulations
MPS450-CO/FET-01 Australia
MPS450-CO/FET-02 Denmark
MPS450-CO/FET-03 India
MPS450-CO/FET-04 Israel
MPS450-CO/FET-05 Italy
MPS450-CO/FET-06 Schuko
MPS450-CO/FET-07 Switzerland
MPS450-CO/FET-08 United Kingdom
MPS450-CO/FET-09 Brazil
Included accessories
BEUNVSL IEC320C14P AC Battery 
Eliminator
9508-0301 Users Manual
This device is not to replace clinical 
testing of waveform detecting devices 
such as patient monitors. The MPS450 
Multiparameter Simulator does not 
provide simulations for all types of fetal 
heart rate tracings and contraction pat-
terns, including the following:
•  variable decelerations
•  sinusoidal pattern
•  reactive tracing
•  variations in FHR variability
• tachysystole
MPS450 optional accessories
HHC3 with MPS450