Specifications
47
Key features
• Handheld
• 12-lead ECG simulation
• 12 arrhythmia selections
• Universal ECG jacks
• Auto sequencing of 
performance waveforms
• Battery operated
• PS410/DPM1B Bundle Kit with 
custom carrying case for quick 
ECG/NIBP patient monitor 
testing
Optional accessories
17191 Carrying Case, single pocket
17192 Carrying Case, double pocket
17024 Universal Banana Adapters
Ordering information
PS410 Patient Simulator
PS410/DPM1B Bundle Kit (includes 
PS410, DPM1B, all accessories, and 
custom carrying case)
Included accessories
CD-ROM PS410 Users Manual 
(electronic, CD)
MANUAL PS410 Users Manual (printed)
BE-UNVSL-IEC320C14 Battery 
Eliminator 100 V ac to 240 V ac
9 V battery
The PS410 ECG/Arrythmia 
Simulator is a compact, high-
performance simulator for patient 
monitor testing.
This handheld device simulates 
a full range of cardiac rhythms 
and a wide variety of ECG con-
ditions. It includes pacemaker 
simulation, 35 arrhythmia selec-
tions, and adult and pediatric 
normal-sinus rhythms.
Small enough to fit in a pocket, 
the handy PS410 weighs less than 
a pound and is easy to operate. 
Technicians simply connect the 
simulator to the device under 
test and use the PS410 keypad to 
enter the code presets. The simu-
lator then transmits the selected 
preset simulations to the device.
Normal sinus rhythm: 12-lead with independent outputs referenced to RL
Normal rate 80 BPM
Selectable rates 30 BPM, 40 BPM, 60 BPM, 80 BPM, 100 BPM, 120 BPM,  
140 BPM, 160 BPM, 180 BPM, 200 BPM, 220 BPM, 240 BPM, 
260 BPM, 280 BPM, and 300 BPM
Accuracy ± 1 %
Output impedance
940 Ω between leads
ECG amplitudes 0.5 mV, 1 mV, and 2 mV
Amplitude accuracy ± 2 % Lead II
High-level output 1000x Lead II
Waveforms Adult or pediatric ECG waveform
ECG performance: Lead II
Square wave 2 Hz and 0.125 Hz
Pulse 30 BPM, 60 BPM, and 120 BPM; 60 ms pulse width
Sine waves 0.5 Hz, 5 Hz, 10 Hz, 40 Hz, 50, and 60 Hz (1 mV amplitude only, 
lead II)
Triangle wave 2 Hz
ST Segment analysis Elevated or depressed: -0.6 mV to 6 mV in 0.2 mV steps
Pacemaker Pacemaker rhythm
Pacer non-capture*
Pacer non-function
Demand pacer with occasional sinus
Demand pacer with frequent sinus
A-V sequential
Artifact selection 50 Hz artifact
60 Hz artifact
Muscle artifact
Baseline artifact
Respiration artifact
*Event occurs once. To repeat, enter the selection again.
PS410
Patient Simulator
Specications
Patient Simulators