Specifications
53
Gas Flow Analyzers
Ordering information
VTMOB/ENG United States, English 
overlay
VTMOB/FRA France, French overlay
VTMOB-8002-02 Denmark, German 
overlay
VTMOB-8002-03 Italy, Italian overlay
VTMOB-8002-04 Spanish, Spanish 
overlay
VTMOB/JPN Japan, Japanese overlay
VTMOB/CHI Chinese, Chinese overlay
VTMOB/BRAZ Brazil, Brazil overlay
Included accessories
VTMOB-7002 Accessory Kit
VTPLUS-WIN CD, includes: quick-
reference card, operators manual, 
getting-started manual, other matter
MANUAL Getting Started Manual
QUICK REF CARD Quick Reference 
Card
VTMOB/HIGH SENSOR High-flow 
Sensor
VTMOB/LOW SENSOR Low-flow 
Sensor
VTMOB/ADAPT FEMALE High-pressure 
Adapter, Male to Female
VTMOB-4402 Low-pressure Adapter
FITTING Oxygen-sensor Cable, 6 ft
VTMOB-4401 T Adapter for Oxygen 
Sensor
97132 Oxygen Sensor
BATTERY 9 V dc Battery (alkaline)
CABLE ASSEMBLY Serial 
Communications Cable (RS-232), 6 ft
CD ROM VT for Windows
®
 PC Software
VTMOB/SOFTCASE Soft Carrying Case
Optional accessories
VTMOB High-pressure Adapter, male 
to male
VTMOB-4405 Temperature and RH 
Sensor, Cable and T Adapter, 6 ft
VTMOB-OWR ADAPTER Power adapter, 
universal (USA and international)
ACCU LUNG ACCU Lung Portable 
Precision Test Lung
High-flow port Low-flow port
Operating flow range ± 200 lpm ± 25 lpm
Accuracy ± 3 % of reading or ± 2 % of range ± 3 % of reading or ± 1 % of 
range
Floor for absolute 
accuracy
25 lpm 3 lpm
Low-flow dropout 2.5 lpm 0.24 lpm
Volume range > ± 60 l ± 60 l
Tidal volume 
accuracy
± 3 % of reading or ± 20 ml, 
whichever is greater
± 3 % of reading or +2 ml, 
whichever is greater
Oxygen measurement Barometric pressure 
measurement
Range 0 % to 100 % 8 psia to 18 psia 
(400 mmHg to 900 mmHg)
Accuracy ± 2 % full-scale output ± 2 % of reading
Sensor technology Galvanic fuel cell —
Calibration Allows user calibration using air 
and 100 % O
2
Not required; however, device 
allows user calibration of offset
Notes:
• Automatic partial pressure compensation for barometric and airway pressure changes.
• Recommended interval for changing oxygen sensor is one year. However, sensor may 
last longer. During user calibration of the sensor, the VT MOBILE can detect if the 
sensor needs to be replaced.
Secondary 
parameter-accuracy 
specifications
Resolution Range Accuracy
Inspiratory and expira-
tory tidal volume
0.1 ml > 10 l ± 3 %
Expiratory minute 
volume
0.001 lpm 0 lpm to 60 lpm ± 3 %
Breath rate 0.1 bpm 2 bpm to 150 bpm ± 1 %
Inspiratory, expira-
tory time
0.01 s 0.25 s to 9.99 s ± 2 % or 0.1 s
Peak inspiratory 
pressure
0.1 cmH
2
O ± 120 cmH
2
O ± 3 % or 1 cmH
2
O
Inspiratory pause 
pressure
0.1 cmH
2
O ± 120 cmH
2
O ± 3 % or 1 cmH
2
O
Mean airway pressure 0.1 cmH
2
O ± 80 cmH
2
O ± 3 % or 1 cmH
2
O
Positive-end expira-
tory pressure (PEEP)
0.1 cmH
2
O -5 cmH
2
O to 
40 cmH
2
O
± 3 % or 1 cmH
2
O
Peak expiratory flow 0.01 lpm 0 lpm to 150 lpm ± 3 % of reading or  
2 % of range
Peak inspiratory flow 0.01 lpm 0 lpm to 150 lpm ± 3 % of reading or  
2 % of range
Temperature 0.1 °C 0 °C to 50 °C ± 1 °C
Units: °C, °F, °K
Humidity 0.1 % 10 % to 95 % ± 10 % RH 
RS-232 serial 
communications
4-pin modular connector located on upper-left side of panel. 
RS-232 compatible with the VT Plus for Windows software 
application (version 2.01.00 or higher.)
Environmental specifications
Operating temperature 10 °C to 40 °C (50 °F to 104 °F)
Storage temperature -25 °C to 50 °C (-13 °F to 122 °F)
Operating humidity 0 % to 80 % non-condensing at temperatures to 31 °C,  
decreasing linearly to 50 % relative humidity at 40 °C (104 °F)
Storage humidity 0 % to 95 % non-condensing
Operating barometric 7 psia to 18 psia
Storage barometric 787.9 mmHg to 522.7 mmHg (-1000 ft to 10000 ft)
VT MOBILE
Portable Gas-Flow Analyzer
Specications