Specifications
54
Key features
• Bidirectional flow, pressure, 
volume, and oxygen 
concentration, and pressure 
measurements
• Low- and high-pressure, and 
flow measurement capability
• Special HF mode—up to  
900 BPM (15 Hz)
• RS-232 and printer ports
• Included Windows-compatible 
graphics software
• All 21 ventilator parameters 
displayed at once on one 
screen
• Operation by user-friendly  
VT PLUS HF command mode or 
special RT-200 command mode
• Minimum, maximum, average, 
absolute, and graph for all 
parameters
• Multiple special-function tests 
for efficient troubleshooting
Optional features
• Operation with a variety of 
precision test lungs available 
from Fluke Biomedical to 
complete a fully NIST-traceable 
ventilator testing system
Optional accessories
5022010 Soft Vinyl Carrying Case for 
VT PLUS HF
9530-0066 Hard-sided Protective 
Carrying Case for VT PLUS HF (limited 
to stock on hand)
Test Lungs
ACCU LUNG ACCU LUNG Portable 
Precision Test Lung (with soft-sided 
carrying case
MI-14900 Michigan Instruments Non-
instrumented Single-adult Test Lung
MI-11000 Michigan Instruments Non-
instrumented Dual-adult Test Lung
MI-12952 Michigan Instruments Non-
instrumented Adult/Infant Test Lung
48499 Siemens 190 Test Lung
48129 Parabolic Airway Resistor ring
Printers
PRINTR/CTZ-UA120V Printer 110 V, 
Citizen IDP 3110
PRINTR/CTZ-US220V Printer 220 V, 
Citizen IDP 3110
71072 Parallel Printer Cable, D25M-C36M
61096 Printer 120 V Power Supply
61097 Printer 220 V Power Supply
97116 DPU-414 and DPU-411 Printer 
Paper (minimum 7 rolls - price is per 
roll)
The VT PLUS HF is Fluke 
Biomedical’s premier 
general-purpose gas flow 
analyzer. In addition, special 
display modes and bidirec-
tional flow make it perfect 
for fully and efficiently 
testing both conventional 
mechanical ventilators and 
high-frequency ventilators. 
EC.6.20 now requires 100 
% completion of scheduled 
life-support device preven-
tive maintenance every year, 
and VT PLUS HF can help 
meet those requirements. 
Multiple special-function tests make troubleshooting quick and efficient.
Learning to use the VT PLUS HF is simple. Technicians control the unit 
using the VT PLUS HF user-friendly command system, or, if they’re familiar 
with the RT-200, they can switch to a special control mode that uses 
RT-200-style commands.
Low-pressure High-pressure Airway-pressure
Range ± 500 mmHg (10 psi) ± 100 psi ± 120 cmH
2
O
Accuracy ± 0.80 % of read-
ing or ± 1.5 mmHg, 
whichever is greater
± 1 % of reading or 
± 0.3 psig, whichever 
is greater
± 0.75 % of read-
ing or ± 0.5 cmH
2
O, 
whichever is greater
Note
Fluid pressure may be 
applied to the positive port; 
however, fluids should 
be kept from entering the 
pressure port by using a 
suitable length of connec-
tion tubing.
Airway pressure is 
internally tapped off the 
proximal-flow sensor port, 
which is the port closest 
to the exhaust port on the 
VT PLUS HF
Low-flow High-flow
Flow range -25 lpm to 25 lpm -300 lpm to 300 lpm
Accuracy ± 1 % of range ± 2 % of range
Low-flow dropout 0.01 lpm —
High-flow dropout — 25 lpm
Volume range  > ± 60 l > ± 60 l
Notes
• 
Tidal-volume accuracy: ± 3 % of reading 
or ± 2 ml, whichever is greater 
• 
Volume accuracy tested to 1 liter
•
 Flow accuracy is specified for dry air or 
oxygen
•
  Below 3.0 lpm, measurement 
accuracy is obtained by allowing the VT 
PLUS HF to fully warm up or manually 
zeroing before reading or documenting 
measurement.
•
 Tidal-volume accuracy: ± 3 % of 
reading or ± 10 ml, whichever is 
greater
•
 Volume accuracy tested to 7 liters
•
  Flow accuracy is specified for dry air 
or oxygen
General
Dimensions (LxWxH) 25.4 cm x 25.4 cm x 12.7 cm (10 in x 10 in x 5 in )
Weight 4.53 kg (10 lb) 
VT PLUS HF
Gas Flow Analyzer
Specications