Specifications
64 Incubator Analyzers
Ordering information
INCU-UNIPOWER Incubator Analyzer
Included accessories
3901000 Users Manual
97156 Soft-Sided Carrying Case
76060 Airflow Sensor
48583FG Universal AC Battery Charger 
with Worldwide Mains Adapter Set
75101FG Serial Cable DB9F to DB9F
48582 Outside Temperature probe 
holder
76058 Adapter for Radiant Infant 
Warmer Assembly
48581 INCU PC Software (one CD)
Key features
• Simultaneous measurement of 
humidity, airflow, sound, and 4 
independent temperatures
• 24-hour continuous testing 
(battery); 35-hour continuous 
testing (main power)
• Battery operated
• Adjustable measurement 
intervals
• Compatible with closed, forced-
convection incubators and open 
infant warmers
• Stand-alone measurement or 
automated testing with PC
• Windows
®
 compatible INCU 
software for easy data 
collection, analysis, and 
documentation
• Numerical and full-color 
graphical reports
Faulty incubator 
controls lengthen 
hospital stays, and 
increase health-
care costs, making 
thorough incubator 
testing essential.
Designed around 
AAMI and IEC stan-
dards that specify 
incubator and radiant 
warmer sound levels 
and thermal char-
acteristics, the INCU 
simultaneously mea-
sures airflow, relative humidity, sound, and four independent temperatures. 
Adjustable measurement intervals allow technicians to configure the unit 
to meet their facilities’ testing requirements. Technicians are free to do 
other work while the device collects and records data.
INCU software lets technicians upload setup parameters and download 
the test results to a PC file, or print the data in reports with full-color 
charts and graphs.
Power supply
Input voltage range 100 V ac to 240 V ac
Input frequency range 47 Hz to 63 Hz
Battery Rechargeable sealed lead-acid type NP7-6 YUASA, 6 V, 7 Ah;  
operates for 24 hours continuously; low-battery alert 
Sound level
Measuring range 30 dbA to 80 dbA
Resolution 0.1 dbA
Accuracy +5 dbA @ 30 dbA to 80 dbA
Relative humidity (without condensation)
Measuring range 0 % to 100 % RH, non-condensing
Resolution 0.1 % RH
Accuracy ± 5 % RH for 0 % to 90 % RH at 77 °F to 104 °F (25 °C to 40 
°C) or 
± 5.3 % RH for 0 % to 100 % RH at 77 °F to 104 °F (25 °C to 40 
°C)
Temperature measurement
Measuring range 5 °C to 70 °C (41 °F to 158 °F)
Resolution 0.1 °C (32 °F)
Accuracy +0.5 °C (+0.9 °F) +1 LSB of range from 25 °C to 40 °C (77 °F to 
104 °F)
Airflow
Measurement range 0.1 m/s to 0.7 m/s
Resolution 0.01 m/s
Accuracy from 0.1 m/s to 0.5 m/s reading ± 0.1 m/s at temperature 
25 °C to 40 °C (77 °F to 104 °F) and humidity 50 % RH ± 15 % 
RH
General information
Measurement interval Via PC: Adjustable from 1 minute to 10 minutes
Storage temperature -20 °C to 50 °C (-4 °F to 122 °F)
Operating temperature 10 °C to 40 °C (50 °F to 104 °F)
Dimensions (LxWxH) 27 cm x 20 cm x 14 cm (10.6 in x 7.8 in x 5.5 in)
Weight 3 kg (6.6 lb)
INCU
™
Incubator Analyzer
Specications