Restore Factory Defaults
If you have saved your own defaults, but would like to return ALL the default settings to those that were set at the factory,
follow the steps below:
4. Touch the SETTINGS icon in the lower right-hand corner of the Operations Screen.
5. Touch “RESTORE FACTORY DEFAULTS.” All of the factory defaults are no restored for all future treatments.
Environmental Conditions
Transport and Storage
is equipment, while packed for transport or storage, should not be exposed to environmental conditions outside the
following ranges:
a. an ambient temperature range of -40°C
b. a relative humidity range of 10% to 100% including condensation
c. an atmospheric pressure range of 500 hPa to 1060 hPa
Operation
is equipment is designed to operate in normal use under the following environmental conditions:
a. an ambient temperature range of +10°C
b. a relative humidity range of 30% to 75% including condensation
c. an atmospheric pressure range of 700 hPa to 1060 hPa
Safety Features of the Dynatron 125
All intensity levels are automatically set to zero at the end of treatment (ensures proper setting of intensity levels for
the next patient).
Internal surge protection protects against line noise, machine switching operation and any other type of interference
that could cause patient discomfort.
e Power Cord is considered the ‘disconnect device’ when it is necessary to ensure that the device is disconnected
from a power source (for service or otherwise). Do not position the device such that it would be dicult to disconnect
the power cord from the device.
Soundhead temperature monitoring prevents the soundhead from becoming too hot, both to protect the soundhead
crystal from damage and to ensure patient comfort.
+70°C
+30°C