Installation Guide User Manual
 EMS Configuration 
Once you are ready to calibrate, select the desired trim that you would like to calibrate, and press 
SEL► to enter its calibration menu. Press the RANGE button to begin calibrating the range of 
the trim. Follow the onscreen instructions, controlling trim to the required position before 
pressing NEXT. Repeat the process for the opposite position. The process will then prompt you 
to put the trim into takeoff position. If you do not require a takeoff indication on the given axis’ 
trim display, you may press NONE. When you have completed the calibration, press the DONE 
button. 
Press the TAKOFF button to calibrate the takeoff position indicator. When calibrated, a green 
line is displayed on the trim scales, indicating takeoff position. 
View the trim display on the EMS Main or Aux page (depending on where you configured it to 
display) to make sure that it works as expected. You may repeat this calibration process as many 
times as you wish. 
Flaps Calibration 
Prior to calibrating your flaps sensor, ensure that you have connected it as 
described on page 3-15 and selected FLAP POS for the desired GP input FUNCT 
parameter as described on page 5-17. 
Enter the EMS menu by pressing any button below an EMS main page and 
pressing MORE > SETUP > MORE > FLAPS. The FLAP CALIBRATION menu 
just shows FLAPS, followed by the GP input that each is configured as (or NONE, 
if no GP input is configured for a given axis). Ensure that this list corresponds to 
the physical connections made during the setup described in the section. Press 
SEL► to enter the flaps calibration menu. 
During the calibration process, ensure that the number shown in the VALUE field changes as 
you adjust the flaps. If the number does not change, the flaps sensor may be incorrectly wired to 
the EMS-D120, or incorrectly configured in the GP input selection under the SENSOR menu. At 
any point in the process, you may press CANCEL to end the calibration without overwriting the 
previous calibration results. 
You must calibrate for at least 2 positions, and may calibrate for as many as 5. The calibration 
process first requires you to put the flaps in the 0° extended position. When you have done this, 
press NEXT. Next, you will calibrate for the second position. Press INC+ or DEC- to set the 
angle that you would like displayed for the second position. If you only wish to have 2 positions 
displayed, press DONE. Otherwise, press NEXT to repeat the process for the third position. 
When you have completed the calibration, press DONE. View the flaps display on the EMS 
Main or Aux page (depending on where you configured it to display) to make sure that it works 
as expected. You may repeat this calibration process as many times as you wish. 
EMS-D120 Installation Guide    5-7 










