Installation Guide Instruction Manual
 Instrument Installation 
FlightDEK-D180 Installation Guide    4-3 
turned on the heading display in the CLUTTR menu). If wiring is not correct, or the EDC-D10A 
is not communicating properly with the FlightDEK-D180, you will see the message REMOTE 
COMPASS NOT DETECTED in place of the heading onscreen. 
The metal shield around the EDC communication cable is connected to the short black/white 
wire emanating from the DB25. Connect this wire to ground. 
Power Inputs 
The FlightDEK-D180 has three separate power inputs, located on the EFIS connector. Of the 
three, only Master Power is required to operate the instrument. The other two inputs provide 
redundancy. Below is a table that explains the three inputs and their purposes. All three of these 
inputs share a common ground signal, wired to pin 3 on the EFIS connector. 
EFIS 
DB25 
pin# 
Function  EFIS 
DB25 
wire 
color 
Description 
1 Master 
Power 
Red  Provides primary power to the instrument. The 
FlightDEK-D180 will switch on upon application of 
power. Connect to a switched power source. Will not 
be adversely affected by engine cranking. 
2 Keep Alive 
Power 
Yellow  A very low current power input which is only used if 
both Master and External Backup Power are not 
applied. Keep Alive draws just enough current to 
keep the clock running. It draws less than 1 milliamp 
of current when not charging the internal backup 
battery. 
Dynon Avionics recommends not connecting the 
Keep Alive wire when the FlightDEK-D180 has an 
internal battery installed or is connected to a GPS. 
Keep Alive is only used in installations with no 
battery or GPS to keep the clock running when the 
unit is powered off. 
3 Master 
Ground 
Black  Connect to ground. Must carry as much as 3 amps. 
Refer to the Grounding section on page 2-2 for more 
detailed information. 
15 External 
Backup 
Power 
(Not wired 
in Dynon 
harness) 
Will operate the FlightDEK-D180 only if Master 
Power is not present. The transition from Master 
Power to External Backup Power will bring up a 
warning, requiring you to press ACK within 30 
seconds to keep the unit operating. This warning will 
also display when transitioning from either Master or 
External Backup to Internal Battery power. 










