PowerSmart Socket POWER QUALITY METER Operation Manual BG0550 Rev.
LIMITED WARRANTY The manufacturer offers the customer a 24-month functional warranty on the instrument for faulty workmanship or parts from date of dispatch from the distributor. In all cases, this warranty is valid for 36 months from the date of production. This warranty is on a return to factory basis. The manufacturer does not accept liability for any damage caused by instrument malfunction.
Do not use the instrument for primary protection functions where failure of the device can cause fire, injury or death. The instrument can only be used for secondary protection if needed. Read this manual thoroughly before connecting the device to the current carrying circuits. During operation of the device, hazardous voltages are present on input terminals. Failure to observe precautions can result in serious or even fatal injury or damage to equipment.
Table of Contents Chapter 1 Introduction ................................................................ 8 Chapter 2 Operating the PowerSmart Socket PQM ................... 14 Meter Assembly ....................................................................................... 14 Controls and Indicators ........................................................................... 15 TEST Mode ...............................................................................................
Setup Menus and Access Rights .............................................................. 50 Viewing and Changing Setup Options ....................................................... 51 Chapter 4 Using Power Software Configuration tool ................ 53 Software Installation ............................................................................... 53 Installing Power Software....................................................................... 53 Installing the USB Driver...................
EN50160 Advanced Setup .................................................................... 129 Clearing EN50160 Evaluation Counters................................................... 130 Configuring Communication Protocols................................................... 131 Configuring Modbus............................................................................. 131 Configuring DNP3 ............................................................................... 132 Configuring Meter Passwords ....
Appendix A Parameters for Monitoring and Logging .............. 189 Appendix B Setpoint Parameters............................................. 207 Appendix C Analog Output Parameters ................................... 208 Appendix D Billing and Load Profile Log Files ......................... 210 Appendix E EN50160 Statistics Log Files ................................ 214 Appendix F EN50160 Evaluation and Recording ..................... 220 EN50160 Background ..................................
Chapter 1 Introduction Chapter 1 Introduction The PowerSmart Socket PQM The PowerSmart Socket PQM is especially designed for utility, industrial and commercial billing metering with high requirements to reliability of power quality monitoring and availability of the device.
Chapter 1 Introduction Meter Features Billing/TOU • Accuracy class 0.2 ANSI C12.20 (current class 10 and 20) active energy and class 0.
Chapter 1 Introduction • High-voltage fast transient recorder; impulsive and low frequency oscillatory transient overvoltages with peaks up to 2kV and durations from 20 microseconds • Ready-for-use compliance supplementary software statistics reports with the Fault Recorder • Digital Fault recorder: three voltage and three/four current channels, up to 10 external digital triggers for monitoring circuit breakers and protection relays; onboard overcurrent and undervoltage fault detector; zero-sequence vo
Chapter 1 Introduction • 8 digital counters for counting pulses from external sources and internal events • 4 programmable timers from 1 cycle to 24 hours for periodic recording and triggering operations on a time basis.
Chapter 1 Introduction Expansion Communication Slot • Serial RS232/RS485 communication port (Modbus RTU/ASCII and DNP3 protocols) • Ethernet port (Modbus/TCP, DNP3/TCP and protocols), up to 5 simultaneous connections IEC 61850 • Wireless GSM/GPRS modem (Modbus/TCP and DNP3/TCP protocols) Backup Power Supply • Auxiliary backup 50-288VAC/90-290VDC power supply module Upgradeable Firmware • New features can be easily added to the meter by simply replacing firmware through any communication port.
Chapter 1 Introduction Meter Nameplate PowerSmart Socket Power Quality Meter 13
Chapter 2 Operating the PowerSmart Socket PQM Meter Assembly Chapter 2 Operating the PowerSmart Socket PQM Meter Assembly The meter view with a removed cover and the location of the expansion slots and lithium battery are shown in the following pictures.
Chapter 2 Operating the PowerSmart Socket PQM Controls and Indicators WATT – watthour pulse LED VAR – varhour pulse LED LCD graphic display Location of the replaceable battery IR optical port TEST button (under cover) DEMAND RESET button SCROLL button SELECT/ENTER button Controls and Indicators Push Buttons The PowerSmart Socket PQM has four push buttons that allow you to control meter operations, scroll through the PowerSmart Socket Power Quality Meter 15
Chapter 2 Operating the PowerSmart Socket PQM TEST Mode multi-page display screens for local meter reading and inspect or change meter setup parameters. See Navigation Buttons in Chapter 3 for buttons location and functionality. Graphical Display Indicators The PowerSmart Socket PQM graphical display is used both for displaying billing and instrumentation data, and for indication of the present tariff rate, time and important diagnostic information.
Chapter 2 Operating the PowerSmart Socket PQM Device Diagnostics kWh resolution. See Energy Pulse LED Indicators above for the test LED pulse rates. You can enter TEST mode directly by an extended press on the TEST button located under the meter cover, or via Power Software. See The TEST Button in Chapter 3 and Device Options and Mode Control in Chapter 5 on how to put your meter in TEST mode and to change the test LED pulse rate. See TEST Mode Data Display for more information on the TEST mode display.
Chapter 2 Operating the PowerSmart Socket PQM Meter Security noise in the region of the device. If the device continuously resets itself, contact your local distributor. Device Fault Alarm The PowerSmart Socket PQM provides a global "DEVICE FAULT" event flag that is asserted all the time while one of the diagnostic events exists. It can be monitored via a setpoint (see Using Control Setpoints in Chapter 5) to give a fault indication through a relay output.
Chapter 2 Operating the PowerSmart Socket PQM Meter Clock and Time Synchronization Password-tampering events are automatically recorded to the Event log file after every three unsuccessful login attempts in 5 minutes. Meter Clock and Time Synchronization Time synchronization provides a common time basis for the meter billing and tariff system, and for the power quality and fault recorders so that events and disturbances can be compared to one another.
Chapter 2 Operating the PowerSmart Socket PQM Billing and Energy Metering • kWh net (Q1+Q4)-(Q2+Q3) - instrumentation • kvarh delivered (Q1+Q2) • kvarh received (Q3+Q4) • kvarh net (Q1+Q2)-(Q3+Q4) - instrumentation • kvarh Q1, Q2, Q3, Q4 quadrants • kVAh delivered (Q1+Q4) • kVAh received (Q2+Q3) • kVAh total (Q1+Q2+Q3+Q4) Energy Measurement Channels The PowerSmart Socket PQM uses two independent current channels for high-accuracy energy measurements and for conventional high-range instrumentation.
Chapter 2 Operating the PowerSmart Socket PQM Billing and Energy Metering Maximum Demand Registers Any of billing energy registers can be individually linked to the maximum demand and cumulative maximum demand registers providing the same demand tariff structure as you selected for energy registers.
Chapter 2 Operating the PowerSmart Socket PQM Instrumentation Metering • 15-min energy load profile for 120 days or 30-min profile for 240 days. Maximum demand profiling for monthly and daily billing data profiles can be individually configured for every register. See Factory Preset Data Log Files in Chapter 5 and Billing and Load Profile Log Files in Appendix D for more information on the file layout and contents.
Chapter 2 Operating the PowerSmart Socket PQM Demand Forgiveness (Cold Load Pickup) Parameter 1/2 cycle 1 cycle 200 ms 1 sec 3 sec 10 sec Flicker Symmetrical components × Phasors × 1 Taken from the energy measurement channel 2 Only for power quality evaluation 3 Not affected by loss compensation 10 min 2 hours × × × Demand Forgiveness (Cold Load Pickup) Demand forgiveness (sometimes called cold load pickup) avoids recording abnormal maximum demands in installations with high pickup cu
Chapter 2 Operating the PowerSmart Socket PQM Power Transformer/Line Loss Compensation Power Transformer/Line Loss Compensation Loss compensation allows accounting for losses in power transformers and/or power line in the event the billing and metering points are located at different sides of the power transformer or at different sides of the power line. The calculated compensation values for kW and kvar losses are added to the measured power quantities and energies.
Chapter 2 Operating the PowerSmart Socket PQM Operating Relay Outputs • Triggering a setpoint – put a digital input into a setpoint trigger list (see Using Control Setpoints in Chapter 5) • Triggering the Fault Recorder – link a digital input to the Fault Recorder (see Configuring Digital Inputs in Chapter 5) and enable external triggers in the Fault Recorder setup (see Configuring the Fault Recorder in Chapter 5) • External synchronization of power demand intervals – select a digital input as an external
Chapter 2 Operating the PowerSmart Socket PQM Operating the Logical Controller All analog outputs are updated at a 1-cycle rate. See Configuring Analog Outputs in Chapter 5 on how to configure analog outputs in your meter.
Chapter 2 Operating the PowerSmart Socket PQM Operating the Fast Transient Recorder Operating the Fast Transient Recorder The fast transient recorder is provided with the optional transient add-on module. It can detect impulsive and low frequency oscillatory transient overvoltages with peaks up to 2kV and durations from 20 microseconds. The recorder does not require any special setup above a conventional PQ recorder setup for transient overvoltages.
Chapter 2 Operating the PowerSmart Socket PQM Communicating with the PowerSmart Socket PQM Optical Port The PowerSmart Socket PQM has an ANSI-compatible optical infrared port for local meter reading via a hand-held unit or a portable PC. The IR port is identified in the PowerSmart Socket PQM as port COM1. It is factory preset to 19200 bps, 8-bits/No-parity data format, and the Modbus RTU protocol.
Chapter 2 Operating the PowerSmart Socket PQM Communicating with the PowerSmart Socket PQM any device address and returns the received address in the response message.
Chapter 2 Operating the PowerSmart Socket PQM Auxiliary AC/DC Backup Power Supply USB Communications An embedded full speed 12-Mbit USB 1.1 port provides a local communications with the support Power Software software. USB communications does not require any settings. Just connect your PC to the PowerSmart Socket PQM USB port using the supplied USB cable and install the USB driver provided on the CD with your meter (see Installing the USB Driver in Chapter 4).
Chapter 3 Using Front Display Display Operations Chapter 3 Using Front Display TEST DEMAND RESET SCROLL SELECT/ENTER Display Operations The PowerSmart Socket PQM has a high-contrast graphical LCD display with backlight for local data read outs, meter setup and servicing. The display operates in two modes: • Multi-page data display mode with Auto-Scroll feature allows you to scroll through display screens and pages to view various billing, instrumentation and status data.
Chapter 3 Using Front Display Display Operations The SCROLL button operates once it’s briefly pressed. It has two functions: • In data display mode, it scrolls through the display pages. • In programming mode, it scrolls through the menu items and allows changing a selected digit when entering numbers. The SELECT/ENTER button normally operates once it’s released.
Chapter 3 Using Front Display Display Operations Navigating in Data Display Mode The following table gives a summary of the button operations in data display mode.
Chapter 3 Using Front Display Display Operations Status Indicators Graphical icons on the bottom status bar give immediate meter status indication and show the present tariff rate. The present date and time are indicated at left on the status bar. The date order can be changed according to local rules via the Local Settings setup.
Chapter 3 Using Front Display Display Operations Indicator Icon Event Priority Description Low General diagnostic message indicator: indicates that there are diagnostic messages you can inspect via the DIAGNOSTICS display. The icon is shut down after you explicitly reset diagnostics messages either from the display or via communications (see Clearing Device Diagnostics). You can disable diagnostic message indicator via the Display Setup menu (see Display Setup).
Chapter 3 Using Front Display Display Operations minutes. You can temporarily set the backlight to continuous operation if you need to work in dark for more time. Auto-Return If the Auto-Return feature is enabled and no button is pressed for the programmable Auto-Return interval (1 to 30 minutes for data displays; fixed at 5 minutes for setup menus), the display automatically returns to the default page from any other data display or a setup menu.
Chapter 3 Using Front Display Data Displays Data Displays The PowerSmart Socket PQM has 12 multi-page data displays listed in the following table. Display Number Display Label Display Contents 1 BILL 0 2 BILL 1 Last (-1) billing period data 3 BILL 2 Second previous (-2) billing period data 4 BILL 3 Third previous (-3) billing period data 5 MAX.
Chapter 3 Using Front Display Data Displays Each billing period display lists all total and tariff energy, maximum demand and cumulative maximum demand registers for all configured billing registers and all active tariffs. The PowerSmart Socket PQM automatically arranges pages in each display.
Chapter 3 Using Front Display Data Displays Register 1 - total readings: kWh delivered kW delivered maximum demand kW delivered cumulative maximum demand Register 2 - total readings: kWh received kW received maximum demand kW received cumulative maximum demand Register 3 - total readings: kvarh delivered kvar delivered maximum demand kvar delivered cumulative maximum demand Register 4 - total readings: kvarh received kvar received maximum demand kvar received cumulative maximum demand Register 1 – tari
Chapter 3 Using Front Display Data Displays Register 1 – tariff 2 readings: kWh delivered kW delivered maximum demand kW delivered cumulative maximum demand Register 2 – tariff 1 readings: kWh received kW received maximum demand kW received cumulative maximum demand … … Register 4 – tariff 2 readings: kvarh received kvar received maximum demand kvar received cumulative maximum demand Previous Billing Period Data The following example demonstrates billing displays for the three previous billing periods.
Chapter 3 Using Front Display Data Displays Register 1 - total readings: kWh delivered kW delivered maximum demand kW delivered cumulative maximum demand Register 2 - total readings: kWh received kW received maximum demand kW received cumulative maximum demand … … Register 1 – tariff 1 readings: kWh delivered kW delivered maximum demand kW delivered cumulative maximum demand … … Engineering Maximum Demand Data Display Maximum demand displays show engineering maximum demands (not billing maximum deman
Chapter 3 Using Front Display Data Displays Delivered kW maximum demand Delivered (inductive) kvar maximum demand Delivered kVA maximum demand Received kW maximum demand Received (capacitive) kvar maximum demand Va-Vc maximum demand Indicate Vab-Vca voltage in line-to-line configurations. Ia-Ic maximum demand Va-Vc THD maximum demand. Indicate Vab-Vca THD in line-to-line configurations.
Chapter 3 Using Front Display Data Displays Ia-Ic THD maximum demand Engineering Data Display Engineering data represent general instrumentation data you can use while installation and inspecting the meter. Use phase angles displays to check the order of phases when connecting wires to the meter terminals. Phase-to-neutral voltages. Only displayed in 4-wire configurations with a neutral.
Chapter 3 Using Front Display Data Displays Total powers and power factor Voltage unbalance, % Current unbalance, % Line frequency Phase voltage THD Indicate phase-to-phase voltage THD in line-to-line configurations Phase current THD and TDD Short-term and long-term voltage flicker 44 PowerSmart Socket Power Quality Meter
Chapter 3 Using Front Display Data Displays Waveform Display The waveform display shows per-phase voltage and current waveforms. Use the SCROLL button to scroll through the phases. Per-phase voltage and current waveforms Harmonics Display Harmonics display shows per-phase voltage and current harmonic spectrum graphs. Use the SCROLL button to scroll through the voltage and current channels.
Chapter 3 Using Front Display Data Displays Phasor Display The phasor display shows a three-phase network phasor diagram. All phase angles are given relatively to the Va channel. Three-phase voltage and current phasor diagram Setup Data Display The setup data display shows basic device settings that can be required for immediate inspecting while meter testing and at the time of installation.
Chapter 3 Using Front Display Data Displays Meter identification info: serial number, production date and calibration date Meter operation time counters (since start of operation): Start of operation date Total operation time in hours Out of service time in hours Total number of power failures Total number of 3-phase voltage interruptions Internal meter temperature Lithium battery status, replacement date and operation time in hours Meter firmware information: Host processor firmware version Fast transie
Chapter 3 Using Front Display Data Displays Device Diagnostics Display The diagnostics display shows device diagnostic messages recorded as a result of the meter self-test diagnostics during start-up and operation. If there are more pages, use a short press on the SCROLL button to scroll through the entire list of messages. Extended press SCROLL + SELECT/ENTER is a shortcut for immediate entering the passwordprotected diagnostics reset menu.
Chapter 3 Using Front Display Programming Mode Programming Mode To enter programming mode from the data display, press and hold the SELECT/ENTER button for more than 5 seconds. Navigation Buttons The following table gives a summary of the button operations in programming mode.
Chapter 3 Using Front Display Programming Mode Password Security The setup menus are secured by 8-digit user passwords. Every time you enter programming mode, you are prompted for a correct password. The meter is primarily shipped with all passwords preset to 9 at the factory. See Meter Security in Chapter 2 for more information on the meter security levels.
Chapter 3 Using Front Display Programming Mode NOTE Access to the Reset menu entries is allowed depending on your security level as shown in Section Reset of Accumulators and Log Files in Chapter 6. If your security level does not allow access to a menu, it will not be listed in the main menu list, and you will not be able to highlight menu items that you are not allowed to change, but you can still view their present settings.
Chapter 3 Using Front Display Programming Mode To exit the submenu and return to the main menu: 1. If the upper-right window is not highlighted yet, highlight it by briefly pressing the SELECT/ENTER button. 2. Press the SELECT/ENTER button for more than 1 second. You will return to the main menu. To exit the main menu and return to the data display: 1. Press briefly the SELECT/ENTER button to highlight the right-upper Exit window. SCROLL 2. Press the SELECT/ENTER button for more than 1 second.
Chapter 4 Using Power Software Configuration tool Software Installation Chapter 4 Using Power Software Configuration tool The support Power Software is a configuration and data acquisition tool that allows you to configure all of the PowerSmart Socket PQM features, monitor your meter online, retrieve recorded files and view reports.
Chapter 4 Using Power Software Configuration tool Software Installation When installation is complete, the Power Software icon appears on your Desktop. Double click on the Power Software icon to run Power Software. For general information on how to work with Power Software, see the “Power Software Getting Started” guide supplied on the installation CD. Installing the USB Driver To take an advantage of the USB communications, install the PowerSmart Socket PQM USB driver on your PC.
Chapter 4 Using Power Software Configuration tool Software Installation 3. Click “Browse”. 4. Point to the “USB” folder located in the Power Software installation directory and click “OK” 5. Click “Next”.
Chapter 4 Using Power Software Configuration tool Creating a New Site for your 6. Click "Finish" to complete installation. The next time you power up the PowerSmart Socket PQM or connect it to your PC with the USB cable, Windows automatically launches the driver for your meter. Creating a New Site for your Meter Power Software keeps all communication and configuration data for your meter in a configuration database called a site database.
Chapter 4 Using Power Software Configuration tool Setting up Communications 3. From the “Look in” box, select the directory where a new database will be stored. By default, it will be the “Sites” directory. Type a site name for your device in the “File name” box, click New, and then click OK. 4. On the Instrument Setup tab, select “PSS” in the “Model” box. Power Software automatically selects the appropriate instrument options for your meter. 5.
Chapter 4 Using Power Software Configuration tool Setting up Communications Communicating through a Serial Port Select Serial Port/Modem Site on the Configuration tab, and then click on the Connection tab to configure your serial port settings. Configuring a Serial Port 1. On the Connection tab, select a COM port from the “Device” box, and then click Configure. 2. Specify the baud rate and data format for the port.
Chapter 4 Using Power Software Configuration tool Setting up Communications connection when using the Modbus RTU or DNP3 protocol. It does not affect Modbus ASCII communications. If there are many applications running on your PC, Power Software might be prevented from responding to received characters fast enough and may close the communication while the device is still transmitting. If you frequently receive the “Communication error” message, try to increase “Break Time-out”.
Chapter 4 Using Power Software Configuration tool Setting up Communications Communicating through a GSM/GPRS Modem You should use Windows Dial-Up networking to access the PowerSmart Socket PQM via your wireless GSM/GPRS modem. See your local GSM/GPRS modem manual and Windows help on how to create a new Dial-Up PPP connection for your GPRS modem. Configuring a Dial-up GPRS Connection To communicate through a GPRS modem: 1. On the Instrument Setup tab, select Internet Site. 2. Click on the Connection tab.
Chapter 4 Using Power Software Configuration tool Setting up Communications 3. Click on the “IP address” and type in the IP address your meter got from the GPRS provider when registering on the GPRS network. You can find the meter GPRS IP address on the GPRS page in the Device Info display (see Device Info Display in Chapter 3) or via the Dial-Up Networking setup dialog in Power Software (see Modem/GPRS IP Address in Setting-Up Dial-Up GPRS Network in Chapter 5) 4.
Chapter 4 Using Power Software Configuration tool Setting Up the Meter Setting Up the Meter Power Software allows you to prepare setup data for your meter off-line without the need to have it connected to your PC. To prepare a setup for your meter: 1. Select the device site from the list box on the Power Software toolbar. 2. Select the desired setup group from the Meter Setup menu. Click on the tab with the setup you want to create or modify. 3.
Chapter 4 Using Power Software Configuration tool Setting Up the Meter Downloading Setups to the Meter You can update each setup in your meter one at a time or download all setups together from the site database. Individual Download To update a particular setup in your meter: 1. Check the On-line button on the Power Software toolbar 2. Select a meter site from the list box on the toolbar. 3. Select the desired setup group from the Meter Setup menu. Click on the setup tab you want to download to the meter.
Chapter 4 Using Power Software Configuration tool Authorization and Security Batch Upload To upload all setups from the device to the site database at once: 1. Check the On-line button on the toolbar. 2. Select the device site from the list box on the toolbar. 3. Select Upload Setups from the Meter Setup menu. NOTE The passwords setup is never uploaded via the Batch Upload and may not be read from the meter. When you open the Passwords setup dialog, all passwords are zeroed.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Communications Chapter 5 Configuring the PowerSmart Socket PQM This chapter describes how to configure the PowerSmart Socket PQM for your particular environment and application from the front display and via Power Software. To access your meter configuration options via Power Software, you should create a site database for your device as shown in Chapter 4.
Chapter 5 Configuring the PowerSmart Socket PQM Display Label Parameter Configuring Communications Options Default Description Protocol Communication protocol Modbus RTU, Modbus ASCII, DNP3, IEC 62056-21 IEC 62056-21 (COM1) Modbus RTU (COM3-4) The communications protocol for the port Interface Port interface RS232, RS485, IR, GPRS IR (COM1) GPRS (COM2) RS485 (COM34) Not changeable; automatically detected by the meter Address Device address Modbus: 1-247 DNP3: 0-65532 IEC 62056-21: 0-65532
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Communications Setting Up the Ethernet Using the Front Display Select Network Setup from the main menu. See Viewing and Changing Setup Options in Chapter 3 for information on navigating in menus. See the table below for available network options. Using Power Software Select Communications Setup from the Meter Setup menu, and then click on the Network Setup tab. The following table lists available network options.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Communications Setting-Up Dial-Up GPRS Network Select Communications Setup from the Meter Setup menu, and then click on the Dial-Up Networking Setup tab. The following table lists available connection options. Parameter Options Default Description Connection Options 68 IP Address 192.168.10.203 The IP address on the dial-in modem PPP network. Not used in a GPRS network. Network Subnet Mask 255.255.0.
Chapter 5 Configuring the PowerSmart Socket PQM Parameter Options Default Modem Initialization AT&F&D1&C1 Phone Number *99# Configuring Communications Description Default modem initialisation string. Do not change. ISP Account The telephone number of the ISP provider. The default number provides a connection to the GPRS network for your PowerSmart Socket PQM GSM/GPRS modem.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Communications The following table lists available options Parameter Options Default Description SNTP Client Enabled NO, YES NO Enables operations of the SNTP client Polling interval, s 60-86400 s 600 s The time remaining requesting time from the SNTP server Primary SNTP Server IP Address 192.36.143.151 The IP address of the primary SNTP server Secondary SNTP Server IP Address 130.149.17.
Chapter 5 Configuring the PowerSmart Socket PQM Parameter Configuring Communications Options Default Description XPW Server Port 0-65535 5001 The TCP service port of the eXpertPower server XPW Client Enabled NO, YES NO Enables operations of the eXpertPower client Time to Next Session, min 1-99999 Connection via Network Ethernet, GPRS/Modem Ethernet Automatically selected depending on the network module installed Time to close the Session, min 1-n 1 The time to keep session open while i
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Communications The server’s IP address, port number and starting Modbus register address are programmable in the meter. The following table lists available options. Parameter Client Enabled Options Default NO, YES Server IP Address Description NO Enables operations of the notification client 192.168.0.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup General Meter Setup Basic Meter Setup The Basic Setup provides the meter with basic information about your electrical network. Using the Front Display Select Basic from the main menu. See Viewing and Changing Setup Options in Chapter 3 for information on navigating in menus. See the table below for available options. Using Power Software Select General Setup from the Meter Setup menu.
Chapter 5 Configuring the PowerSmart Socket PQM Display Label Parameter CT Primary CT Primary Current CT4 Primary General Meter Setup Options 1 Default Description 1-20,000 A 5A The primary rating of the phase current transformer I4 CT Primary Current 1-20,000 A 5A The primary rating of the I4 current transformer Nom. Frequency Nominal Frequency 50, 60 Hz 60 Hz The nominal line frequency Power Dmd.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Device Options and Mode Control Using the Front Display To enter the device options setup menu, select Options from the main menu. To enter the TEST mode menu, select Test Mode Setup from the main menu. You can also directly enter the TEST menu by extended press on the TEST button located under the meter cover (see The TEST Button in Chapter 3). See Viewing and Changing Setup Options in Chapter 3 for information on navigating in menus.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup The following table lists available device options. Display Label Parameter Options Default Description Power/Energy Options Power Mode Power Calculation Mode S=f(P, Q) (reactive S=f(P, Q) power mode), Q=f(S, P) (nonactive power mode) The method used for calculating reactive and apparent powers (see “Power Calculation Modes” below) Energy Roll Energy Roll Value, kWh 1000.0 10000.0 100000.0 1000000.0 10000000.0 100000000.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup 1. When the reactive power calculation mode is selected, active and reactive powers are directly measured and apparent power is calculated as: 2. S = P2 + Q2 3. This mode is recommended for electrical networks with low harmonic distortion, commonly with THD < 5% for volts, and THD < 10% for currents. In networks with high harmonics, the following method is preferable. 4.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup 7 Open Closed Closed 8 Closed Closed Closed Instrument Transformer Correction Setup Transformer correction allows you to compensate ratio and phase angle inaccuracies of the voltage and current instrument transformers. Select General Setup from the Meter Setup menu, and then click on the Transformer Correction tab. The available options are listed in the following table.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup current. When a single test point is defined, it is applied over the full operating range of the transformer. The order in which you define the test points does not matter. The meter will automatically arrange all data in a correct order. Check the Correction Enabled box to activate transformer correction in the meter. You can also enable and disable transformer correction from the front display (see Device Options and Mode Control).
Chapter 5 Configuring the PowerSmart Socket PQM Parameter Billing Point Location Options Supply side, far end, Supply side, transformer end, Load side, transformer end, Load side, far end Number of Metering Elements 2 (Open Delta) 3 (WYE) General Meter Setup Description Defines the location of the billing point: far end – near the supply source or near the load, transformer end – near the power transformer. Defines the number of metering elements in the meter depending on the wiring mode.
Chapter 5 Configuring the PowerSmart Socket PQM Billing Point Supply side, far end Metering Point Supply side, far end General Meter Setup Supply side, transformer end Load side, transformer end -Supply line losses -Supply line losses -Supply line losses -Transformer losses -Transformer losses -Load line losses -Transformer losses -Transformer losses -Load line losses Supply side, transformer end +Supply line losses Load side, transformer end +Supply line losses +Transformer losses +Transformer
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Local Settings This setup allows you to specify your local time zone, daylight savings options and a time synchronization source. Using the Front Display Select Local from the main menu. See Viewing and Changing Setup Options in Chapter 3 for information on navigating in menus. See the table below for available options. Using Power Software Select General Setup from the Meter Setup menu, and then click on the Local Settings tab.
Chapter 5 Configuring the PowerSmart Socket PQM Display Label Parameter Options General Meter Setup Default Description Daylight Daylight saving time (DST) Disabled Enabled Scheduled Disabled Disabled: the RTC operates in standard time only. Enabled: the time is automatically updated at the predefined fixed DST switch dates. Scheduled: the time is automatically updated at the calendar scheduled DST switch dates. Start Month DST Start Month JanuaryDecember March The month when DST begins.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Using the IRIG-B Select the GPS IRIG-B option and connect the GPS master clock to the meter IRIG-B input. See the PowerSmart Socket PQM Installation Manual for the connection diagram. The PowerSmart Socket PQM automatically synchronizes its clock with the GPS time each second, normally with accuracy better than one millisecond if the time is locked to the GPS satellite time.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Display Setup Using the Front Display Select Display from the main menu. See Viewing and Changing Setup Options in Chapter 3 for information on navigating in menus. See the table below for available display options. Using Power Software Select General Setup from the Meter Setup menu, and then click on the Display Setup tab. The Display Setup consists of Displays Options parameters and Custom Displays setup menus.
Chapter 5 Configuring the PowerSmart Socket PQM Display Label 86 Parameter Options General Meter Setup Default Description Auto-Scroll Auto-Scroll Interval Disabled, 2-10, 15, 20, 25, 30 seconds 5 Defines the display scroll interval or disables autoscroll Scroll Sequence Auto-Scroll Sequence 1 1-2 1-3 1-4 1-5 1-6 1-6, 10 1-6, 10-11 1-6, 10-12 1, 5, 10-12 1-2, 5, 10-12 1-3, 5, 10-12 1, 6, 10-12 1-2, 6, 10-12 1-3, 6, 10-12 1-4, 6, 10-12 1, 10-12 1-2, 10-12 1-3, 10-12 1-4, 10-12 1, 11-12 1-2, 11-1
Chapter 5 Configuring the PowerSmart Socket PQM Display Label Front Panel Access mode Parameter Front Panel Security Options Full access, View only General Meter Setup Default Full access Description At full access mode, the user is able to use buttons to navigate and/or setup device meter parameters.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Up to three custom displays numbered 13, 14 and 15 (usually NORMAL, ALTERNATE and TEST) with up to 8 custom pages in each are available. The customer can select whatever he wants to be displayed on the NORMAL display – Display Title, and what to be displayed to other display number.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Using the Browse tab, select the Logo in bmp format to be uploaded, and then click OK.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Configuring Digital Inputs The PowerSmart Socket PQM has two embedded digital inputs (DI1-DI2) and can provide one additional expansion module with a total of 8 digital inputs (DI3 through DI10). I/O numbers are automatically assigned to the inputs. To configure the digital inputs, select General Setup from the Meter Setup menu, and then click on the Digital Inputs tab. Digital inputs that are not present in your meter are not shown.
Chapter 5 Configuring the PowerSmart Socket PQM Parameter Fault Recorder Options Checked Unchecked Default Unchecked General Meter Setup Description When the box is checked, a positive transition on the digital input (open to closed transition event) triggers the Fault recorder. Debounce Time The debounce time is assigned in groups of 2 adjacent digital inputs. DI1 and DI2 have the same debounce time, while DI3 and DI4 are allowed to have another setting.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Configuring Relay Outputs The PowerSmart Socket PQM has one embedded relay output (RO1) and can provide one additional expansion module with a total of 6 relay outputs (RO2 through RO7). I/O numbers are automatically assigned to outputs. To configure the relay outputs in your meter, select General Setup from the Meter Setup menu, and then click on the Relay Outputs tab. Relays that are not present in your meter are not shown.
Chapter 5 Configuring the PowerSmart Socket PQM Parameter Options Default General Meter Setup Description Polarity NORMAL (N.O.) INVERTING (N.C.) NORMAL Normal polarity: the relay is normally deenergized in its non-active state and is energized in its active (operated) state. Inverting polarity: the relay is normally energized in its non-active state and is deenergized in its active (operated) state. It is called failsafe relay operation.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Configuring Analog Outputs The PowerSmart Socket PQM can provide one AO expansion module with a total of 4 analog outputs with options for 0-1 mA, ±1mA, 0-20 mA and 4-20 mA. To configure the analog outputs, select General Setup from the Meter Setup menu, and then click on the Analog Outputs tab. The setup entries will be blanked if the AO expansion module is not installed in the meter.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup parameter actually covers a lower range, you can change the scales to provide a better resolution on an analog output. Scaling Non-Directional Analog Outputs For non-directional analog outputs with a 0-1mA, 0-20mA or 4-20mA current option, you can change both zero and full engineering scales for any parameter. The engineering scale need not be symmetrical.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup • ±2 mA: set the 1 mA scale to ½ of the required full-scale output for both uni-directional and bi-directional parameters. For example, to provide the 0 to 2 mA output current range for Volts measured by the meter in the range of 0 to 120V, set the 1 mA scale to 60V; then the 120V reading will be scaled to 2 mA.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Using Counters The PowerSmart Socket PQM provides 8 universal nine-digit counters that you can use for counting different events. Each counter can be independently linked to any digital input and count input pulses with a programmable scale factor. You can link a number of digital inputs to the same counter. Each counter can also be incremented through the Control Setpoints in response to any internal or external event.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Using Periodic Timers The PowerSmart Socket PQM has 4 programmable interval timers that are normally used for periodic recording and triggering operations on a time basis through the Control Setpoints. Whenever a programmed timer interval is expired, the timer generates an internal event that can trigger any setpoint (see Using Control Setpoints). The programmable time interval can be from 1 cycle and up to 24 hours.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Using Control Setpoints The PowerSmart Socket PQM has an embedded logical controller that runs different actions in response to userdefined internal and external events. Unlike a PLC, the PowerSmart Socket PQM uses a simplified programming technique based on setpoints that allows the user to define a logical expression based on measured analog and digital values that produces a required action.
Chapter 5 Configuring the PowerSmart Socket PQM Option Relation Format/Range <=, >=, =, ON, OFF General Meter Setup Description The relational operator used in the conditional expression for the trigger Operate limit The threshold (in primary units) at which the conditional expression would be evaluated to true. Not applicable for digital triggers. Release limit The threshold (in primary units) at which the conditional expression would be evaluated to false.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup Using Numeric Triggers For numeric (analog) triggers, a setpoint allows you to specify two thresholds for each trigger to provide hysteresis (dead band) for setpoint operations. The Operate Limit defines the operating threshold, and the second Release Limit defines the release threshold for the trigger. The trigger thresholds are always specified in primary units.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup setpoint triggers. When the setpoint event is asserted, the timer is restarted, and then generates the next timer event when the timer interval expires. If you wish to record interval data at predefined intervals without linking to other events, just select a timer as a setpoint trigger and specify in the setpoint actions list a data log file you want to use for recording.
Chapter 5 Configuring the PowerSmart Socket PQM General Meter Setup logging and waveform logging that are shared between all setpoints using an OR scheme for each separate target. A relay output is operated when one of the setpoints linked to the relay is activated and stays in the operated state until all of these setpoints are released (except for latched relays that require a separate release command to be deactivated).
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Billing/TOU Configuring Billing/TOU The PowerSmart Socket PQM provides 10 universal billing energy registers that can be linked to any internal energy source or to any external pulse source that delivers pulses through the device digital inputs. Any of energy registers can provide either a single-tariff energy accumulation or be individually linked to the TOU system providing both total and multi-tariff energy billing.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Billing/TOU Configuring Billing/Tariff Registers To configure the meter billing/TOU Energy/TOU from the Meter Setup menu.
Chapter 5 Configuring the PowerSmart Socket PQM Parameter Options Configuring Billing/TOU Default Description Channel None, Ch1-Ch6 None Defines a measurement channel for multimeter energy metering. Not used for a single self-metering channel if there are no external sources. Set to Ch1 for internal energy sources, and to Ch2 through Ch6 for external sources in the event of multichannel energy metering.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Billing/TOU To configure your daily profiles: 1. Select the desired season and day type. 2. Select the start time for each tariff change point and the corresponding active tariff number. 3. Repeat the setup for all active profiles. The first tariff change point is fixed at 00:00 hours, and the last tariff change you specified will be in use until 00:00 hours on the next day.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders settings and to select a proper daily tariff schedule for any day within a year. The above picture gives you an example of a single-season tariff schedule configured for weekends and the designated U.S.A. holidays. To configure your season tariff schedule: 1. In the “Season/Period” box, select the season, and in the “Day Type” box, select a day type for this calendar entry. 2.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders The meter memory is fully configurable except of the files that keep monthly and daily billing profiles and energy load profile. They are pre-configured at the factory and may not be changed by the user. Two of the data log files are automatically configured in your meter for recording EN51060 compliance statistics data and harmonics survey data.
Chapter 5 Configuring the PowerSmart Socket PQM Option Parameters Range Configuring Recorders Description 0-16 The number of parameters in a single data record The device memory can be partitioned for a total of 22 log files: • Event log • 16 Data logs • 3 Waveform logs • EN50160 PQ event log • Fault event log Memory is allocated for each file statically when you set up your files and will not change unless you re-organize the files. To change the file properties or to create a new file: 1.
Chapter 5 Configuring the PowerSmart Socket PQM File Configuring Recorders Record Size, Bytes File Size, Bytes 1068 × Number of Channels Waveform Log EN50160 Power Quality Log 32 Record size x Number of series (events) × Number of records per series Record size × Number of records Fault Log 40 Record size × Number of records The meter memory is pre-configured at the factory for common data trending, billing, power quality and fault recording applications as shown in the following table. No.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders 2. Select a desired file type for your file. 3. Select the maximum number of records you want to be recorded in the file. 4. Click OK, then send your new setup to the meter or save to the device database. By default, the Event recorder stores all events related to configuration changes, resets, and device diagnostics. In addition, it can record events related to setpoint operations.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders 2. Select a partition type for your file. 3. Select the number of parameters you want to be recorded in the file records. 4. Select the maximum number of records you want to be recorded in the file. 5. Click OK, and then send your new setup to the meter, or save to the device database. To define the contents of the file: 1.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders neighboring parameters: when you open the “Group” box for the next parameter, Power Software highlights the same group as in your previous selection; if you select this group again, Power Software automatically updates the “Parameter” box with the following parameter in the group. 4. Add the name for your data log file in the “Name” box. It will appear in the data log reports. 5.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders Data Log #2 No.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders Data Log #14 No. Parameter 1 Generic voltage V1/V12 2 Generic voltage V2/V23 3 Generic voltage V3/V31 4 Generic current I1 5 Generic current I2 6 Generic current I3 7 Generic current I4 8 Generic frequency EN50160 Statistics Data Logs Data log #9 and Data log #10 are pre-configured for recording EN50160 compliance statistics and harmonics survey data.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders Configuring the Waveform Recorder The PowerSmart Socket PQM can store waveforms in three log files. Configurable waveform log files #1 and #2 can record waveforms at four programmable sampling rates: 32, 64, 128 or 256 samples per cycle. A log file stores up to 7 channels simultaneously (three voltage channels and four current channels) with a programmable length and up to 20 pre-event cycles.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders 2. Select a file type for your file. 3. Select the number of channels for simultaneous recording in the file to define the file record size. 4. Select the maximum number of records you want to be recorded in the file. 5. The number of records required to store a single waveform series per event is defined as follows: Number of Records per Series = Sampling Rate (Samples per Cycle) x Number of Cycles per Event / 512 6.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders The following table lists available waveform options. Parameter Recording Time Options Fixed Event-controlled Description Fixed – a waveform is recorded up to the maximum specified length Event-controlled – a waveform extends for as long as trigger conditions exist plus post-event time, or up to the maximum specified length Samples per Cycle 32, 64, 128, 256, 1024 Waveform sampling rate Max.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders 5. Save your waveform setup to the device database, and then send it to the meter.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders Configuring the Fault Recorder The Fault recorder automatically records all fault events to the Fault log file. It can be triggered via the embedded fault detector, or externally through any of the 8 digital inputs. The Fault recorder can be globally disabled or enabled in your meter.
Chapter 5 Configuring the PowerSmart Socket PQM Option Range Default Configuring Recorders Description Hysteresis, % 0-50.0% 5.0 Defines the hysteresis for a fault trigger in percent of the threshold Trigger Enabled Checked Unchecked Checked Links a fault trigger to the Fault Recorder. If the box is unchecked, the Fault Recorder does not respond to the trigger. On Start Checked Unchecked Checked On End Checked Unchecked Unchecked Enables waveform log when the fault event ends Log No.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Recorders adjust the thresholds according to your network conditions before enabling triggers. The Current Unbalance trigger has a different calculation algorithm than the common current unbalance measurements. Since the unbalance readings give a relation of the maximum deviation from the average to the phase average current, the value could produce high readings for low currents.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring the EN50160 Configuring the EN50160 Recorders See EN50160 Evaluation and Recording in Appendix F for information on EN50160 evaluation techniques. Basic Device Settings The following device settings affect the EN50160 evaluation and should be checked prior to running the EN50160 recorders.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring the EN50160 Advanced Setup. The evaluation period for harmonics survey can be selected independently from the EN50160 compliance evaluation. You can upload and view the harmonics survey data collected by your device via Power Software reports or via common Power Software data logs views.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring the EN50160 To configure the PQ recorder: 1. Select Memory/Log from the Meter Setup menu, and then click on the EN50160 PQ Recorder tab. 2. Adjust thresholds and hysteresis for PQ triggers if required. The harmonic and interharmonic voltage limits can be individually set for each harmonic order via the EN50160 Harmonics Setup. Limits for the signaling voltage frequencies are automatically taken from the “Meister-curve”. 3.
Chapter 5 Configuring the PowerSmart Socket PQM Option Range Configuring the EN50160 Default Description PQ Log Threshold, % 0-200.0% Defines the operating threshold for a PQ trigger in percent of the nominal (reference) value Hysteresis, % 0-50.0% Enabled Checked Unchecked On Start Checked Unchecked Checked Enables waveform log when a PQ event starts On End Checked Unchecked Unchecked Enables waveform log when a PQ event ends Log No. 1-2 5.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring the EN50160 conventional waveforms to Waveform log #1 or #2, then a second longer waveform will be also recorded, synchronized with the fast transient waveform. You can temporary disable the PQ recorder in your meter. To enable or disable the PQ recorder: 1. Check or uncheck the Recorder Enabled checkbox. 2. Send your setting to the device.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring the EN50160 The default EN50160 compliance limits are shown in the picture above. You can change the number of the evaluated harmonics and interharmonics via the EN50160 Advanced Setup. EN50160 Advanced Setup The EN50160 Advanced Setup allows you to configure the EN50160 evaluation options in your meter. To configure the EN50160 evaluation options: 1. Select Memory/Log from the Meter Setup menu, and then click on the EN50160 Advanced Setup tab.
Chapter 5 Configuring the PowerSmart Socket PQM Option Range Configuring the EN50160 Default Description Rapid Voltage Changes Repetition Rate 1-10 1 Defines the maximum repetition rate in variations per hour (equal or less than) for rapid voltage changes. Voltage changes at higher rates are not classified since they will be subject for flicker. Pst Period 1-10 min 10 min THD, up to order 25-50 40 Defines the highest harmonic order included in the THD evaluation.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Communication Configuring Communication Protocols This section describes how to customize protocol options for use with your application software. Configuring Modbus Modbus Point Mapping The PowerSmart Socket PQM provides 120 user assignable registers in the address range of 0 to 119.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Communication 3. Type in the actual addresses you want to read from or write to via the assignable registers. Refer to the PowerSmart Socket PQM Modbus Reference Guide for a list of the available registers. Note that 32-bit Modbus registers should always start at an even register address. 4. Click Send to download your setup to the device. Configuring DNP3 DNP Options can be changed both via DNP3 and Modbus.
Chapter 5 Configuring the PowerSmart Socket PQM Parameter Options Binary Input Change Event Object Without Time With Time Number of AI to Generate events 0-64 Analog Input Object Analog Input Change Event Object Configuring Communication Default With Time Description The default BI change event object variation for requests with qualifier code 06 when no specific variation is requested Analog Inputs (AI) 3 43 The total number of AI change event points for monitoring 32-bit 32-bit –Flag 16-bit
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Communication 3 The total number of AI, BI and BC change event points may not exceed 64. When you change the number of the change event points in the device, all event setpoints are set to defaults (see Configuring DNP Event Classes below).
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Communication The factory-set Class 0 point ranges are shown in the picture below. Configuring DNP Event Classes The PowerSmart Socket PQM generates object change events for any static analog input, binary input, and binary counter point when a corresponding point either exceeds a predefined threshold, or the point status changes. A total of 64 change event points are available for monitoring.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Communication The number of event setpoints for each static object type is specified via the DNP Options setup. NOTE The device clears all event buffers and links the default set of static points to each event object type every time you change the number of points for any of the objects. To define setpoints for selected static points: 1. Check the “Ext” box if you wish to use the extended point list. 2.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Communication returns below the threshold minus hysteresis – applicable for AI objects; a predefined return • Less than (under) - a new event is generated when the point value drops below the specified threshold, and then when it returns above the threshold plus a predefined return hysteresis – applicable for AI objects. Hysteresis for the return threshold is 0.05 Hz for frequency and 2% of the operating threshold for all other points. 4.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Meter Passwords Configuring Meter Passwords Using the Front Display Select Access from the main menu. You should have administration rights to enter this menu. See Viewing and Changing Setup Options in Chapter 3 for information on navigating in menus. The setup menu allows you to configure three passwords for three security levels the meter provides. To setup the meter passwords: 1. Use the SCROLL passwords. button to scroll through the 2.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Meter Passwords The present passwords settings are never uploaded from the meter via the Passwords setup. When you open the dialog, all passwords are zeroed. To setup new passwords: 1. Enter the first (lower-level) password in the “Password 1” box and repeat it in the following “Confirm password” box. 2. In the same manner, enter Password 2 and Password 3 for the medium and high-level security passwords. 3.
Chapter 5 Configuring the PowerSmart Socket PQM Configuring Meter Passwords 6. Enter the new password and repeat it in the following “Confirm new password” box. 7. Click Send to update the password in the meter.
Chapter 6 Device Control and Upgrading Updating the Meter Clock Chapter 6 Device Control and Upgrading This chapter describes how to perform control functions in your meter from the front panel display and via Power Software. To access device control options from Power Software, you should have your meter online and provide a correct password with respective permissions. Updating the Meter Clock Using the Front Display To enter the clock setup menu, select RTC Setup from the main menu.
Chapter 6 Device Control and Upgrading Clearing Device Diagnostics You need not update the clock in your meter if the clock is synchronized to the external GPS master clock. Clearing Device Diagnostics Using the Front Display See the following section Reset of Accumulators and Log Files on how to clear the device diagnostics from the front display. Using Power Software Via Power Software, you can both examine the present device diagnostics status, and clear it.
Chapter 6 Device Control and Upgrading Reset of Accumulators and Log Files Access to the Reset menu entries is allowed depending on your security level as shown in the following table. Clearing files is not allowed from the front display. Reset Entry Menu Function Security Level Diagnostic Reset of device diagnostics Low Power MD Volt/Amp MD Volt MD Ampere MD Harmonic MD All MD Reset of engineering maximum demands Low Device Oper.
Chapter 6 Device Control and Upgrading Reset of Accumulators and Log Files Using Power Software Select Reset from the Monitor menu. Refer to the table above for required permissions. To clear log files, you should have the administration rights. To reset the accumulation counters or to clear a file: 1. Click on the corresponding button. If a reset entry has more than one target, you are allowed to select targets from the dialog box. 2. Check the corresponding boxes, and then click OK. 3.
Chapter 6 Device Control and Upgrading Reset of Billing Maximum Demands NOTE The “Clear All Log Files” entry does not affect billing data and energy load profile files. Reset of Billing Maximum Demands A demand reset signals an end of the present billing period and causes the following automatic actions: • increments the number of the recorded billing periods.
Chapter 6 Device Control and Upgrading Master Reset (Clearing Billing Data) 1. Select Reset from the Monitor menu. 2. Click on the Reset Billing/TOU Max. Demands button. 3. Confirm your command. Master Reset (Clearing Billing Data) The PowerSmart Socket PQM permits the clearing billing data via the Master Reset menu. This menu also allows you to reset the meter configuration setups to their factory defaults.
Chapter 6 Device Control and Upgrading Operating Event Flags Operating Event Flags The PowerSmart Socket PQM has 8 common event flags that are intended for use as temporary event storage and can be tested and operated from the control setpoints. You can transfer an event to the setpoint and trigger its operation remotely by changing the event status through Power Software. To enter the Event Flags dialog, select Device Control from the Monitor menu, and then click on the Event Flags tab.
Chapter 6 Device Control and Upgrading Upgrading Meter Firmware To change the status of an event flag: 1. In the “Status” box, select the desired flag status. 2. Click on Send. Upgrading Meter Firmware You can upgrade device firmware through any communication port installed in your meter: a serial port, USB, wireless GSM/GPRS modem, or via the Internet. Downloading firmware is only supported through the Modbus RTU/ASCII and Modbus/TCP protocols.
Chapter 6 Device Control and Upgrading Upgrading Meter Firmware 2. Point to the firmware upgrade file, click Open, and then confirm upgrading the meter. 3. When asked for the password, password, and click OK. type the meter 4. Wait until Power Software completes downloading the file. It takes about 9-10 minutes at 115,200 bps via a serial port, or about 20 seconds via a USB port, to download the file to the meter. 5.
Chapter 6 Device Control and Upgrading Upgrading Meter Firmware logon again so the IP address you used for connecting to the meter will no longer be supported. You should check the meter for a new GPRS network IP address either from the front display, or via Power Software using another communication port.
Chapter 7 Monitoring Meters Viewing Real-time Data Chapter 7 Monitoring Meters Viewing Real-time Data Real-time data is continuously retrieved from your devices and updated on the screen at the rate you defined in the Instrument Setup. To get real-time data from your device: 1. Ensure that the On-line button on the Power Software toolbar is checked. 2. Select the device site from the list box on the Power Software toolbar. 3.
Chapter 7 Monitoring Meters Viewing Real-time Data Polling Devices To run data polling, click on either the Poll button Continuous Poll button Stop button or on the local toolbar. Click on the to stop continuous polling. You can open as many monitor windows as you wish, either for different sites, or for the same site using different data sets. An open data monitor window is linked to the current site and does not change if you select another site in the site list.
Chapter 7 Monitoring Meters Viewing Real-time Data Viewing a Data Table Changing the Data View Power Software displays data in either a single record or multi-record view. To change the view, click on the Data Monitor window with the right mouse button and select either Wrap to see a single record, or UnWrap to go to the multirecord view.
Chapter 7 Monitoring Meters Viewing Real-time Min/Max Log Printing Data To check the report, as it will look when printed, select Print Preview from the File menu. button on the Power To print retrieved data, click on the Software toolbar, select a printer, and then click OK. Real-time Data Logging Power Software allows you to store data records to a database automatically at the time it updates data on the screen. To setup the real-time logging options: 1. Open the Data Monitor window. 2.
Chapter 7 Monitoring Meters Viewing Real-time Harmonic Spectrum See Working with Tables in Chapter 9 for more information on working with tables. Viewing Real-time Harmonic Spectrum To retrieve real-time harmonic spectrum from your meter: 1. Check the On-line button toolbar. on the Power Software 2. Select the device site from the list box on the toolbar. 3. Select RT Harmonic Monitor from the Monitor menu.
Chapter 8 Retrieving and Storing Files Uploading Files on Demand Chapter 8 Retrieving and Storing Files Using Power Software, you can retrieve recorded events, data and waveforms from your devices and save them to files on your PC in the Microsoft Access database format. Historical data can be uploaded on demand any time you need it, or periodically through the Upload Scheduler that retrieves data automatically on a predefined schedule, for example, daily, weekly or monthly.
Chapter 8 Retrieving and Using the Upload Scheduler Storing Files 6. If you wish to retrieve data starting with a known date, check the “From” box and select the start date for retrieving data. 7. If you wish to retrieve data recorded before a known date, check the “To” box and select the last date for retrieving data. 8. Click OK. Using the Upload Scheduler To setup the Upload Scheduler: 1. Select Upload Scheduler from the Logs menu.
Chapter 8 Retrieving and Storing Files Using the Upload Scheduler 2. Click Add Site, point to the site database for which you want to organize the schedule, and then click OK. 3. Click Browse and select a database for storing retrieved data, or type the name for a new database, select a directory where you want to save it, and then click OK. 4. Click Configure or double click on the site row. 5. Select a daily, weekly or monthly schedule, and adjust the start time.
Chapter 8 Retrieving and Retrieving EN50160 Statistics Files Storing Files does not cause loss of data, since the scheduler will resume operations when you check this button again. Suspending the Scheduler To suspend the Upload Scheduler, check the Suspend Scheduler box at right. To activate the Upload Scheduler, leave this box unchecked. Running the Scheduler on Demand You can run the scheduler at any time outside the schedule by checking the Start Now box at right. This is a one-time action.
Chapter 8 Retrieving and Storing Files Viewing Files On-line See Viewing EN50160 Statistics Reports on how to get the EN50160 compliance report for the latest online statistics stored in the database. Viewing Files On-line Sometimes, it is useful to review a particular piece of historical data on-line at the time you expect new events to appear in the log. Power Software allows you to retrieve historical data from a particular log without storing it to a file.
Chapter 8 Retrieving and Exporting Files Storing Files 1. Click on the Export toolbar. button on the Power Software 2. Select the database and a waveform or data log table you want to export, and then click Open. 3. Select a directory where you want to store your exported files, type a file name that identifies your files, select a desired file output format, and then click on the Save button. The PQDIF files are commonly recorded in compressed format.
Chapter 8 Retrieving and Storing Files Exporting Files Automatic Converting Power Software allows you to automatically convert waveform and data logs into COMTRADE or PQDIF format at the time you upload data from your devices via the Upload Scheduler. To automatically convert your waveform or data log tables into COMTRADE or PQDIF format: 1. Open the Upload Scheduler. 2. Highlight a desired device site with the left mouse button, and then click on the Export button. 3.
Chapter 8 Retrieving and Archiving Files Storing Files Exporting Files in Excel Format Power Software can automatically convert data files into the Microsoft Excel workbook format while retrieving data from your meters via the Upload Scheduler. To store files in Excel format, follow instructions in the previous section and select Excel Workbook as the output file format.
Chapter 8 Retrieving and Storing Files Archiving Files 3. Check the Enable box and select a periodic schedule for archiving your files for this site. 4. Click OK. To avoid archiving partially updated data, archiving is performed in a day after expiring a scheduled period and not before 2 hours a.m.
Chapter 9 Viewing Files and Reports Operations with Files Chapter 9 Viewing Files and Reports Operations with Files Files you read from the meters are stored in one or in a number of tables in the meter database. Sections of multisection files like waveforms, load profiles and PQ statistics files are stored in multiple tables – each file section in a separate database table. Opening a Database Table To open a database table: 1. Click on the Open button on the Power Software toolbar, or select “Open...
Chapter 9 Viewing Files and Reports Viewing Options does not support this format, you may instruct Power Software to drop milliseconds. To change the way Power Software records and displays the timestamp: 1. Select Options from the Tools menu and click on the Preferences tab. 2. Select the preferred timestamp format. 3. Click OK. Working with Tables Selecting Font and Grid To change the table font or a type of the grid lines: 1.
Chapter 9 Viewing Files and Reports Viewing Options To print a table to a printer, click on the print button the toolbar, select a printer and click OK. on Working with Graphic Windows Selecting Channels To select the channels you want to view on the screen, click on the graph window with the right mouse button, select “Channels...”, check the channels you want displayed, and then click OK. Checkboxes for channels that are not available in the present view are dimmed.
Chapter 9 Viewing Files and Reports Viewing Options The minimum distance between the two markers is exactly one cycle. To change the marker position, click on the button, or click on the window with the right mouse button and select Set Marker, and then click on the point where you want to put the marker. You can also drag both markers with the mouse, or use the right and left arrow keys on your keyboard to change the marker position.
Chapter 9 Viewing Files and Reports Viewing Options Printing a Graph To check how the graph appears on a printed page, select Print Preview from the File menu. To print a graph to a printer, click on the print button the toolbar, select a printer and click OK.
Chapter 9 Viewing Files and Reports Viewing the Event Log Viewing the Event Log The Event log contains time-tagged events related to configuration changes, resets, device diagnostics, and setpoint operations. The Event log is displayed in a tabular view, one event per row. Use the scroll bar to view the entire log contents. See Working with Tables for more information on viewing options.
Chapter 9 Viewing Files and Reports Viewing the Event Log Linking to Waveforms and Data Records If a setpoint is programmed for logging setpoint operation events and it could trigger waveform or data recording, Power Software automatically establishes links between the event and other database records where it finds a relationship with the event. The event ID for which Power Software found related data is blue colored. Click on the colored ID to check a list of the event links.
Chapter 9 Viewing Files and Reports Viewing the Power Quality Event Log Viewing the Power Quality Event Log The EN 50160 PQ event log stores individual time-tagged power quality events. The PQ log file is displayed in a tabular view, one event per row. Use the scroll bar to view the entire log contents. See Working with Tables for more information on viewing options.
Chapter 9 Viewing Files and Reports Viewing the Power Quality Event Log Linking to Waveforms and Data Records PQ events for which Power Software found related links are blue colored. Click on the colored ID to check a list of the event links. Click on a list item to move to the related waveform or data log records. Data records associated with the event are taken into a separate window for easy viewing and trending.
Chapter 9 Viewing Files and Reports Viewing the Power Quality Event Log See Using a Zoom and Changing Region Size for information on how to change the graph scale to separate closely located elements.
Chapter 9 Viewing Files and Reports Viewing the Fault Log Viewing the Fault Log The Fault recorder stores time-tagged fault events. Fault log files are displayed in a tabular view. Use the scroll bar to see the entire log contents. See Working with Tables for more information on viewing options. Filtering and Sorting Events To filter events, click on the Filter button , or click on the report window with the right mouse button and select “Filter...
Chapter 9 Viewing Files and Reports Viewing the Fault Log Linking to Waveforms and Data Records When displaying the fault report, Power Software establishes links between the event and related waveforms and data log records. Fault events for which Power Software finds related links are blue colored. Click on the colored event ID with the left mouse button to check a list of the event links. Click on a list item to move to the related waveform or data log records.
Chapter 9 Viewing Files and Reports Viewing the Data Log Viewing the Data Log Data log files can be displayed in a tabular view, one data record per row, or in a graphical view as a data trend graph. Viewing Data Trend To view data in a graphical form, click on the Data Trend button on the local toolbar. To change the time range for your graph, click on the Time Range button on the local toolbar, and then select the desired date and time range.
Chapter 9 Viewing Files and Reports Viewing Waveforms Viewing Waveforms Waveforms can be displayed in five different views: as overlapped waveforms on a common time axis, individual waveforms using separate time axes, as a cycle-by-cycle plot of RMS values, as a frequency chart for voltage channels, and as a harmonic spectrum in a graph or in a table view. Each waveform window has a local toolbar from where you can open another window to examine the waveforms in a different view.
Chapter 9 Viewing Files and Reports Viewing Waveforms A waveform window displays up to 128 waveform cycles. If the waveform contains more cycles, the scroll bar appears under the waveform pane allowing you to scroll through the entire waveform. Scrolling through the Database The status bar at the bottom shows you how many records the database contains. Use green arrowheads to scroll through records.
Chapter 9 Viewing Files and Reports Viewing Waveforms Viewing a Frequency Plot Click on the button to view a cycle-by-cycle frequency plot for the sampled voltage waveforms. Viewing a Spectrum Chart button to view a spectrum chart for the Click on the selected waveform channel. Power Software provides voltage, current, active power and reactive power spectrum charts. To change a channel, click on the window with the right mouse button, select “Channels...
Chapter 9 Viewing Files and Reports Viewing Waveforms To review or change harmonic limits: 1. Click on the spectrum window with the right mouse button and select “Limits…”. 2. Select a harmonics standard, or select “Custom” and specify your own harmonic limits.
Chapter 9 Viewing Files and Reports Viewing Waveforms 3. Check the Enabled box to visualize harmonic faults on the spectrum graph and in harmonic tables. Harmonics that exceed selected compliance levels are colored in red on the graph and in the tables. Viewing a Spectrum Table Click on the button on the local toolbar to display the harmonics spectrum in a tabular view for a selected phase or for all phases together.
Chapter 9 Viewing Files and Reports Viewing Waveforms If you leave the Triangle box checked, Power Software connects the ends of the voltage and current vectors showing you three-phase voltage and current triangles. This is useful when analyzing voltage and current unbalances. Phasor diagrams are calculated over one waveform cycle pointed to by the left marker line. As you move the marker, the phasor diagrams are updated reflecting the new marker position.
Chapter 9 Viewing Files and Reports Viewing Waveforms 2. Open a waveform you want to synchronize with other waveforms, and then click on the Multi-site View button . Power Software searches for timecoordinated waveforms that have the same time span as your selected waveform. 3. Check the sites your want to see displayed. 4. Click on the “Channels” button and select channels for each site. 5. Click OK. To change the channels, click on the waveform window with the right mouse button and select “Channels..
Chapter 9 Viewing Files and Reports Viewing EN50160 Statistics Reports Viewing EN50160 Statistics Reports Viewing the EN50160 Compliance Report To get the EN50160 Compliance report, select “EN50160 Compliance Statistics” from the Reports menu, point to the database where you stored the retrieved statistics data, uncheck the voltage characteristics’ tables you do not want to be reported, and then click Open. An example of the EN50160 compliance report is shown in the following picture.
Chapter 9 Viewing Files and Reports Viewing EN50160 Statistics Reports Selecting the Time Range To change the time range or contents of the report, click on the report with the right mouse button, select “Options…”, select the required time range, check the voltage characteristics to be included in the report, and then click OK.
Chapter 9 Viewing Files and Reports Viewing EN50160 Statistics Reports Customizing Reports If you wish to add a logo image, header and footer to your reports: 1. Select “Report Setup…” from the Reports menu, or click on the report window with the right mouse button, and then select “Report Setup…”. 2. Click on the Change button and select a logo image file. Check the “Show” box to include your logo into a report. 3. Type the header text in the Page Header box.
Chapter 9 Viewing Files and Reports Viewing EN50160 Statistics Reports 5. Click OK. Both the header and the footer may contain more than one line of the text. Use the Enter button to move to the next line as usually. Viewing the EN50160 Online Statistics Report If you retrieved the EN50160 online statistics data, you can view the online report on the last retrieved statistics in the same manner as the EN50160 Compliance statistics report.
Appendix A Parameters for Monitoring and Logging Appendix A Logging Parameters for Monitoring and The following table lists all electricity and status parameters available for monitoring via communication ports, for data logging, and for triggering setpoints.
Appendix A Parameters for Monitoring and Logging Designation 190 Description MINUTES Minutes SECONDS Seconds MINUTE INTERVAL Minute interval: 1-5, 10, 15, 20, 30, 60 min SYMM COMP Symmetrical Components V PSEQ Positive-sequence voltage V NSEQ Negative-sequence voltage V ZSEQ Zero-sequence voltage V NSEQ UNB% Negative-sequence voltage unbalance V ZSEQ UNB% Zero-sequence voltage unbalance I PSEQ Positive-sequence current I NSEQ Negative-sequence current I ZSEQ Zero-sequence current
Appendix A Parameters for Monitoring and Logging Designation Description 8 PF L3 Power factor L3 V1 THD V1/V12 voltage THD 2,3 V2 THD V2/V23 voltage THD 2,3 V3 THD V3/V31 voltage THD 2,3 I1 THD I1 current THD 3 I2 THD I2 current THD 3 I3 THD I3 current THD 3 I1 KF I1 K-Factor 3 I2 KF I2 K-Factor 3 I3 KF I3 K-Factor 3 I1 TDD I1 current TDD 3 I2 TDD I2 current TDD 3 I3 TDD I3 current TDD 3 V12 V12 voltage V23 V23 voltage V31 V31 voltage RT LOW 1-Cycle Low Values
Appendix A Parameters for Monitoring and Logging Designation 192 Description 8 PF LEAD Total PF lead kW IMP Total kW import (delivered) kW EXP Total kW export (received) kvar IMP Total kvar import (delivered) 8 8 8 8 kvar EXP Total kvar export (received) V AVG 3-phase average line-to-neutral/line-to-line voltage V LL AVG 3-phase average line-to-line voltage I AVG 3-phase average current RT AUX 1-Cycle Auxiliary Values I4 I4 current In In current FREQ Frequency V UNB% Voltage u
Appendix A Parameters for Monitoring and Logging Designation V2x Description V2x phase-to-ground voltage V3x V3x phase-to-ground voltage AVR LOW 1-Second Low Values on any Phase V LOW Low line-to-neutral/line-to-line voltage I LOW Low current kW LOW Low kW kvar LOW Low kvar kVA LOW Low kVA PF LAG LOW Low lagging PF PF LEAD LOW Low leading PF THD LOW Low voltage THD 2,43 V THD LOW Low current THD 4 KF LOW Low K-Factor I TDD LOW Low current TDD 1 4 4 V L-L LOW Low line-to-line
Appendix A Parameters for Monitoring and Logging Designation 194 Description V UNB% Voltage unbalance 6 I UNB% Current unbalance 6 V4x V4x neutral-ground voltage Temp Internal temperature RMS (0.2 sec) 0.
Appendix A Parameters for Monitoring and Logging Designation Description I2 I2 current I3 I3 current I4 I4 current In In current V ZERO-SEQ Zero-sequence voltage I ZERO-SEQ Zero-sequence current V UNB% Negative-sequence voltage unbalance I UNB% Negative-sequence current unbalance FREQ Frequency V PSEQ Positive-sequence voltage V ZSEQ UNB% Zero-sequence voltage unbalance HRM TOT (0.2 sec) 0.
Appendix A Parameters for Monitoring and Logging Designation 196 Description I2 TDD I2 TDD I3 TDD I3 TDD I4 TDD I4 TDD I1 KF I1 K-Factor I2 KF I2 K-Factor I3 KF I3 K-Factor I4 KF I4 K-Factor HRM TOT (10 min) 10-Minute Total Harmonics V1 THD V1/V12 THD 2 V2 THD V2/V23 THD 2 V3 THD V3/V31 THD 2 I1 THD I1 THD I2 THD I2 THD I3 THD I3 THD I4 THD I4 THD V1 THD/I V1/V12 interharmonics THD 2 V2 THD/I V2/V23 interharmonics THD 2 V3 THD/I V3/V31 interharmonics THD 2 I1 THD
Appendix A Parameters for Monitoring and Logging Designation Description I2 DMD I2 Ampere demand I3 DMD I3 Ampere demand kW IMP BD kW import (delivered) block demand kvar IMP BD kvar import (delivered) block demand kVA BD kVA block demand kW IMP SD kW import (delivered) sliding window demand kvar IMP SD kvar import (delivered) sliding window demand kVA SD kVA sliding window demand kW IMP ACD kW import (delivered) accumulated demand kvar IMP ACD kvar import (delivered) accumulated demand
Appendix A Parameters for Monitoring and Logging Designation ENERGY 198 Description Instrumentation Total Energy kWh IMPORT kWh delivered kWh EXPORT kWh received kWh NET kWh net kvarh IMPORT kvarh delivered kvarh EXPORT kvarh received kvarh NET kvarh net kvarh Q1 kvarh quadrant Q1 kvarh Q2 kvarh quadrant Q2 kvarh Q3 kvarh quadrant Q3 kvarh Q4 kvarh quadrant Q4 kVAh TOTAL kVAh total kVAh IMPORT kVAh delivered (Q1+Q4) kVAh EXPORT kVAh received (Q2+Q3) SUMMARY REGS Billing Summary
Appendix A Parameters for Monitoring and Logging Designation I3 %HD01 Description H01 Harmonic distortion I3 %HD02 H02 Harmonic distortion … ... I3 %HD50 H50 Harmonic distortion ANG V1 V1/V12 Harmonic Angles V1 H01 ANG H01 Harmonic angle V1 H02 ANG H02 Harmonic angle … ... V1 H50 ANG H50 Harmonic angle ANG V2 V2/V23 Harmonic Angles V2 H01 ANG H01 Harmonic angle V2 H02 ANG H02 Harmonic angle … ...
Appendix A Parameters for Monitoring and Logging Designation 200 Description kvar L2 H01 kvar L2 8 kvar L3 H01 kvar L3 8 kVA L1 H01 kVA L1 8 kVA L2 H01 kVA L2 8 kVA L3 H01 kVA L3 8 PF L1 H01 Power factor L1 8 PF L2 H01 Power factor L2 8 PF L3 H01 Power factor L3 8 HRM TOT POW Fundamental Total Power Values kW H01 Total fundamental kW kvar H01 Total fundamental kvar 8 8 8 kVA H01 Total fundamental kVA PF H01 Total fundamental PF FLICKER Flicker V1 Pst V1/V12 short-ter
Appendix A Parameters for Monitoring and Logging Designation Description I1 TDD MIN I1 current TDD I2 TDD MIN I2 current TDD I3 TDD MIN I3 current TDD V12 MIN V12 voltage V23 MIN V23 voltage V31 MIN V31 voltage MIN TOTAL Minimum 1-Cycle Total Values kW MIN Total kW kvar MIN Total kvar 8 8 8 kVA MIN Total kVA PF MIN Total PF PF LAG MIN Total PF lag PF LEAD MIN Total PF lead MIN AUX Minimum 1-Cycle Auxiliary Values 8 8 8 I4 MIN I4 current In MIN In current FREQ MIN Frequen
Appendix A Parameters for Monitoring and Logging Designation 202 Description I3 TDD MAX I3 current TDD V12 MAX V12 voltage V23 MAX V23 voltage V31 MAX V31 voltage MAX TOTAL Maximum 1-Cycle Total Values kW MAX Total kW kvar MAX Total kvar 8 8 8 kVA MAX Total kVA PF MAX Total PF PF LAG MAX Total PF lag PF LEAD MAX Total PF lead MAX AUX Maximum 1-Cycle Auxiliary Values 8 I4 MAX I4 current In MAX In current 8 8 FREQ MAX Frequency V UNB% MAX Voltage unbalance 6 I UNB% MAX
Appendix A Parameters for Monitoring and Logging Designation Description REG10 CMD Summary register #10 cumulative maximum demand TOU PRMS TOU Parameters ACTIVE TARIFF Active TOU tariff ACTIVE PROFILE Active TOU profile TOU REG1 Billing TOU Energy Register #1 REG1 TRF1 Tariff #1 register REG1 TRF2 Tariff #2 register … … REG1 TRF8 Tariff #8 register TOU REG2 Billing TOU Energy Register #2 REG2 TRF1 Tariff #1 register REG2 TRF2 Tariff #2 register … … REG2 TRF8 Tariff #8 register
Appendix A Parameters for Monitoring and Logging Designation 204 Description REG10 TRF2 Tariff #2 register … … REG10 TRF8 Tariff #8 register TOU MAX DMD REG1 Billing TOU Maximum Demand Register #1 REG1 TRF1 MD Tariff #1 maximum demand REG1 TRF2 MD Tariff #2 maximum demand … … REG1 TRF8 MD Tariff #8 maximum demand REG1 TRF1 CMD Tariff #1 cumulative maximum demand REG1 TRF2 CMD Tariff #2 cumulative maximum demand … … REG1 TRF8 CMD Tariff #8 cumulative maximum demand TOU MAX DMD REG2
Appendix A Parameters for Monitoring and Logging Designation Description … … REG6 TRF8 MD Tariff #8 maximum demand REG6 TRF1 CMD Tariff #1 cumulative maximum demand REG6 TRF2 CMD Tariff #2 cumulative maximum demand … … REG6 TRF8 CMD Tariff #8 cumulative maximum demand TOU MAX DMD REG7 Billing TOU Maximum Demand Register #7 REG7 TRF1 MD Tariff #1 maximum demand REG7 TRF2 MD Tariff #2 maximum demand … … REG7 TRF8 MD Tariff #8 maximum demand REG7 TRF1 CMD Tariff #1 cumulative maximum d
Appendix A Parameters for Monitoring and Logging 6 The value is calculated using a simplified technique as a relation of the maximum deviation of phase values from a 3phase average value to a 3-phase average. 7 The value is calculated using a simplified technique as a one third of the RMS value of a 3-phase vector sum. 8 Not affected by loss compensation – always indicate readings at the metering point.
Appendix B Setpoint Parameters Appendix B Setpoint Parameters Setpoint Triggers For setpoint triggers, see Monitoring and Logging. Appendix A Parameters for Setpoint Actions The available setpoint actions are listed in the following table.
Appendix C Analog Output Parameters Appendix C Analog Output Parameters The following table lists parameters that can be provided on the meter analog outputs.
Appendix C Analog Output Parameters Designation Description 1-Sec Auxiliary Values In AVR In Current FREQ AVR Frequency Present Demands kW IMP ACD Accumulated kW import demand kW EXP ACD Accumulated kW export demand kvar IMP ACD Accumulated kvar import demand kvar EXP ACD Accumulated kvar export demand kVA ACD Accumulated kVA demand 1 In 4LN3, 4LL3, 3LN3 and 3LL3 wiring modes, the voltages are line-to-neutral; for any other wiring mode, they are line-to-line voltages.
Appendix D Billing and Load Profile Log Files Appendix D Billing and Load Profile Log Files The following tables show the record structure for the monthly and daily billing data profile and energy load profile log files. The second column shows data abbreviations used in the Power Software data log reports. Data log file sections are highlighted in bold. Energy Load Profile Data Log (Data Log #12) Field No.
Appendix D Billing and Load Profile Log Files Field No. Designation Description Monthly Maximum Demand Register #1 1 REG1 MD Summary (total) max. demand 2 TIME0 Summary (total) max. demand timestamp 3 REG1 CMD Summary (total) cumulative max. demand 4 TRF1 MD Tariff #1 max. demand 5 TIME1 Tariff #1 max. demand timestamp 6 TRF1 CMD Tariff #1 cumulative max. demand 7 TRF2 MD Tariff #2 max. demand 8 TIME2 Tariff #2 max. demand timestamp 9 TRF2 CMD Tariff #2 cumulative max.
Appendix D Billing and Load Profile Log Files Field No. Designation Description 22 TRF7 MD Tariff #7 max. demand 23 TIME7 Tariff #7 max. demand timestamp 24 TRF7 CMD Tariff #7 cumulative max. demand 25 TRF8 MD Tariff #8 max. demand 26 TIME8 Tariff #8 max. demand timestamp 27 TRF8 CMD Tariff #8 cumulative max.
Appendix D Billing and Load Profile Log Files Field No. Designation Description 6 TRF5 MD Tariff #5 max. demand reading 7 TRF6 MD Tariff #6 max. demand reading 8 TRF7 MD Tariff #7 max. demand reading 9 TRF8 MD Tariff #8 max. demand reading The number of parameters in each section is automatically configured depending on the number of actual tariffs you defined in the TOU Daily Profiles.
Appendix E EN50160 Statistics Log Files Appendix E EN50160 Statistics Log Files The following tables list the EN50160 evaluation parameters recorded by the device to the EN50160 statistics data log files. The second column shows data abbreviations used in the Power Software data log reports. Data log file sections are highlighted in bold. EN50160 Compliance Statistics Log (Data Log #9) Field No.
Appendix E EN50160 Statistics Log Files Field No.
Appendix E EN50160 Statistics Log Files Field No.
Appendix E EN50160 Statistics Log Files Field No.
Appendix E EN50160 Statistics Log Files Field No.
Appendix E EN50160 Statistics Log Files Field No.
Appendix F EN50160 Evaluation and Recording Appendix F EN50160 Evaluation and Recording EN50160 Background The EN50160 European standard “Voltage characteristics of electricity supplied by public distribution systems” issued by CENELEC defines the main physical characteristics of electric energy supplied by low and medium voltage public distribution systems under normal operating conditions. The voltage characteristics are evaluated using a statistical approach.
Appendix F EN50160 Evaluation and Recording values, which are intended to provide users with information on the order of magnitude which can be expected. The following table gives the characteristics for which indicative values have been specified by the standard. Voltage Characteristic Indicative values Observation period Voltage dips Less than 1 s, 60% depth Year Short interruptions 70% less than 1 s Year Long interruptions 10 to 50% less than 3 min Year Temporary overvoltages Less than 1.
Appendix F EN50160 Evaluation and Recording clears the evaluation registers and counters so that each evaluation period’s statistics is stored in a separate record. You can upload and view the online statistics data via Power Software reports, using data collected since the beginning of the present evaluation interval. You can also manually clear the present contents of the counters though Power Software before starting your EN50160 evaluation.
Appendix F EN50160 Evaluation and Recording Supply Voltage Variations This characteristic defines slow variations of steady state supply voltage magnitude. Method of Evaluation The basic supply voltage magnitude measurement is the RMS value of the steady state voltage over a period of 10 minutes under normal operating conditions. A voltage variation is not evaluated if the supply voltage crosses a voltage tolerance limit (±15% Un).
Appendix F EN50160 Evaluation and Recording the supply voltage. Flicker is indicated by the long-term flicker severity parameter Plt, which is evaluated every 2 hours. Method of Evaluation The basic measurement is the short-term flicker severity indicator Pst, evaluated each 10 minutes by instrumentation complying with IEC 61000-4-15. The indicative long-term flicker severity Plt is evaluated from 12 consecutive Pst values.
Appendix F EN50160 Evaluation and Recording Indicative Values Under normal operating conditions the expected number of voltage dips in a year may be from up to a few tens to up to one thousand. The majority of voltage dips have a duration less than 1 s and a depth less than 60%. Voltage Interruptions Voltage interruptions correspond to temporary loss of supply voltage on all phases lasting less than or equal to 3 minutes in the event of short interruptions, and more than 3 minutes for long interruptions.
Appendix F EN50160 Evaluation and Recording becomes lower than the end threshold on all affected phases including a threshold hysteresis. The overvoltage threshold can be changed in the meter via the EN50160 PQ Recorder setup. The basic voltage measurement is one-cycle RMS voltage updated each half-cycle.
Appendix F EN50160 Evaluation and Recording Voltage Unbalance This characteristic defines the magnitude and/or phase asymmetries of three-phase steady state supply voltage. Method of Evaluation The basic measurement is the RMS value of the steady state voltage unbalance over a period of 10 minutes under normal operating conditions. It is defined using the theory of symmetrical components by the negative sequence component expressed in percent of the positive sequence component.
Appendix F EN50160 Evaluation and Recording Interharmonic Voltage Method of Evaluation Since the EN50160 does not specify target limits for interharmonic voltages, this feature is normally disabled in your meter. You can enable evaluation of interharmonic voltages via the EN50160 Advanced Setup. The basic measurements are the individual interharmonic voltage distortion factors (HD) and the total interharmonic distortion factor (THD) over a period of 10 minutes under normal operating conditions.
Appendix F EN50160 Evaluation and Recording evaluation of signaling voltages via the EN50160 Advanced Setup. The PowerSmart Socket PQM evaluates up to four ripple control frequencies. You can select the required signaling frequencies via the EN50160 Advanced Setup. The basic measurement is the magnitude of the signaling voltage over a period of 3 seconds under normal operating conditions. Signaling voltages are not evaluated if the supply voltage crosses a voltage tolerance limit (±15% Un).
Appendix G Data Scales Appendix G Data Scales The maximum values for volts, amps and powers in the PowerSmart Socket PQM setup and in communications are limited by the voltage and current scale settings. See Device Options and Mode Control in Chapter 5 on how to change the voltage and current scales in your meter. The following table shows the meter data scales.
Appendix H Device Diagnostic Codes Appendix H Diagnostic Code Device Diagnostic Codes Display Message Description Reason 2 RAM/Data fault RAM/Data error Hardware failure 3 Watchdog reset Hardware watchdog reset Hardware failure 4 Sampling fault DSP/Sampling fault Hardware failure 5 CPU exception CPU exception Hardware failure 6 Run-time error Run-time error Hardware failure 7 SW watchdog reset Software watchdog timeout Hardware failure 8 Power down Power down Loss of power
Appendix H Device Diagnostic Codes 232 PowerSmart Socket Power Quality Meter