Manual

EAGLE MICROSYSTEMS. INC.
Instrumentation for Measurement and Control
4 copyright 2/02
ALARM SCREEN:
Press the pushbutton under CRT or DGR, to configure. Press ESC to return.
1 CRT Critical alarm screen to set value for relay contact.
2 DGR Danger alarm screen to set value for relay contact.
3
4 ESC Go back to the monitor setup screen.
CRITICAL ALARM SETUP SCREEN:
From the critical alarm setup screen you can view the current value. To
change, press the pushbutton under UP or DWN to change the value. Press
the pushbutton under SET to enter the new value and return back to alarm
setup screen.
1UPTo increase the value.
2 DWN To decrease the value.
3
4 SET Go back to the alarm setup screen.
DANGER ALARM SETUP SCREEN:
From the danger alarm setup screen you can view the current value. To
change, press the pushbutton under UP or DWN to change the value. Press
the pushbutton under SET to enter the new value and return back to alarm
setup screen
1UPTo increase the value.
2 DWN To decrease the value.
3
4 SET Go back to the alarm setup screen.
Premier Series Monitor
CALIBRATION:
SELECT CALIBRATION
From the MONITOR SETUP SCREEN, press the pushbutton under CAL.
1 CAL Enter calibration screen..
2 ALM Enter alarm select screen.
3
4 RUN Go back to the main run screen
CALIBRATION SCREEN:
Press the pushbutton under CH1 or IOUT. Press ESC to return.
1 CH1 Press to calibrate 4-20mA analog input..
2
3 IOUT Press to calibrate 4-20mA analog output.
4 ESC Will ask to set new password yes /no
SEL SO2 CH1 ALM
CRT DGR ESC
CRITICAL 10.0
UP DWN SET
DANGER 2.0
UP DWN SET
MONITOR SETUP
CAL
ALM RUN
CALIBRATION
CH1
IOUT ESC