Instruction manual

3 - 5
☼☼● : 1.8 L / minute
☼☼●☼ : 1.95 L / minute maximum value
☼☼☼☼: 2.25 L / minute or more
(4) Warning message
When water flow is over the maximum limit, E-13 error will be shown on the Evacuation control
panel. No buzzer sounds in the case.
If it is less than the lower limit, error E-12 will be shown and buzzer sound, and a message
indicating (Objective lens over heat or water is not supplied) will be shown on the GUI. These will
be gone off when water flow is adjusted.
3.1.3 Starting the Display
(1) Turn the DISPLAY switch to 1 (ON).
(2) Windows 2000 will start up and request to key in [Ctrl + Alt +
Delete].
Then the logon dialog window will appear. Enter login name and
password. Initially no password is set for Administrator.
The user may set or change password (refer to the manual of
Windows 2000).
NOTICE
Use S-4800 on the Administrator level.
(4) The S-4800 system will start up automatically and the initial log-in dialog window appears.
(To start the S-4800 system after once closing it, select PC_SEM from start menu)
Input login name and its password. The SEM main window will open.
Refer to <3.9.16 Password Setting and 3.9.17 Setting Login Name>