Instruction manual

1-5
Figure 1-1 shows the relation between frequency of floor vibration and vibration transmittance of
the instrument. Since the vibration transmittance is maximum at the resonance point (about
2 Hz) of the anti-vibration mount provided with the instrument, the allowable amplitude is
minimum at this point.
NOTICE: Please consult with us if the floor vibration exceeds the allowable values. (Vibration
should be measured in advance if it is expected to cause a disturbance.)
Low frequency
vibration range
Frequency : fn
5 Hz
Resonance point
Transmittance Tr
Fig. 1-1 Vibration Transmittance - Frequency Characteristics