Instruction manual

2 - 4
Fig. 2-4 shows the Type 1 specimen stage.
Fig. 2-4 Specimen Stage (type I)
[X] control : Moves specimen in X direction within a range from 0 to 50 mm.
[Y] control : Moves specimen in Y direction within a range from 0 to 50 mm.
[Z] control : Moves specimen in Z direction or shifts WD (Working Distance)
within a range from 1.5 to 40 mm.
(X-ray analysis position is at 15 mm working distance)
[R] control : Rotates specimen through 360° (continuously variable)
[T] control : Tilts specimen within a range from -5° to 70°.
[Absorbed current measuring terminal] : Measures beam absorbed current on specimen.
(Normally used as specimen applied voltage
terminal)
[Opening selector] : Selects any of 4 aperture openings.
[Fine opening control X knob] : Effects fine control in the direction for introducing or
taking out aperture.
[Fine opening control Y knob] : Effects fine control in the direction orthogonal to the
above direction.
[Stage lock switch] : Used for locking the specimen stage to reduce
vibration influence.
Fine opening control X knob
Opening selector
Fine opening control Y knob
[T] control
Stage lock switch
[X] control
[Z] control
[R] control
Absorbed
current
measuring
terminal
[Y] control