2 - 9
Fig. 2-8b shows the Display / Operation area.
Fig. 2-8a Display / Operation area
Vacuum of each chamber, status of evacuation sequence and messages are displayed on this
area.
• Status indicator lamps :
When each lamp (IP1 to SEC) is lit, it shows that the corresponding chamber is in good
vacuum.
Blinking lamp shows that the Numeric / character display is indicating vacuum of the
corresponding chamber. For example, when the SC lamp is blinking and the displayed
value is 3E-3, it indicates that vacuum of specimen chamber is 3x10-3 Pa.
• Display mode selection switches :
MODE : By pressing the switch, blinking status indicator lamp will move to next, in the
order of IP1ÆIP2ÆIP3ÆS.CÆS.E.CÆECOÆBAKEÆInitial (No indication).
SELECT : When the above status is in Initial, ECO and BAKE, the SELECT switch further
changes the Numeric / character display to lower level indication.
• Operation switches :
UP DOWN switch
Increase or decrease time setting for ECO mode and BAKE mode
START STOP switch
Start and stop baking and ECO mode.
• Numeric / character display
Indicates status code, vacuum, time and messages.
Numeric / character
display
Status indicator lam
p
s
Display mode
selection switches
Operation
switches