Instruction manual
2 - 12
[Alarm according to gun baking]
Message
No.
description recovering operation
E31(+
buzzer)
HV cable head is not placed to t
the cable holder properly.
E32(+
buzzer)
Baking protection cover is not put.
E34(+
buzzer)
Ion pump high voltage is not
applied.
E35(+
buzzer)
Ion pump 1 vacuum degraded
while baking
E36(+
buzzer)
Baking time setting is not correct.
Refer to 4.10 Gun baking.
[Alarm according to ECO mode setting]
E37(+
buzzer)
Start gun baking while in ECO
mode
E38(+
buzzer)
Evacuation start time must be 5
hours or more later from present
time.
E39(+
buzzer)
Start ECO mode while gun baking
E99(+
buzzer)
Specimen exchange rod is not at
its proper position or at present in
other errors (E01, E02, E03,
E05, E08, E09, E06, E14)
Refer to ECO mode operation
[Errors due to equipment failure]
E01(*)
Control power supply error
(DC24V)
E02(*)
Control power supply error
(DC15V)
E03(*)
Control power supply error
(DC12)
E04(*)
Vacuum valve did not work
correct
E05(*)
Rotary pump (or dry pump) is not
working
E06(*)
GV did not work correct
E07(*)
Filament of Pi gauge broken
Contact to service engineer