Cleaning the Platen
The platen is the inside bottom surface of the exposure module. Over time, this
surface can collect dust. This dust can create artifacts on printed images.
Supplies Required
3M Auto-Pak Tack Cloth (60-9800-0888-6)
Procedure
1. From the local panel, press Supply to close the film cartridge and open the
supply door.
2. Remove the film cartridge.
!
Warning
When the power cord is plugged in, hazardous voltages are present in some areas
of the imager. These voltages can cause severe injury or death.
3. Turn off the imager and unplug its power cord.
4. Open the left door via its mechanical release (see page 4-14).
5. Open the platen access door (turn the handle counterclockwise and push in the
door).
!
Caution
The anti-reflective surface of the platen is easily damaged by fingernails and jewelry.
Before cleaning the platen, remove any jewelry (rings, bracelets, watches, etc.)
which may accidentally come in contact with the platen.
If the Auto-Pak cloth catches on parts inside the exposure module, take care not to
damage the parts when removing the cloth. Check for and remove any torn pieces of
cloth.
6. Use an Auto-Pak cloth to wipe the surface of the platen, as well as the inside
surfaces close to the bottom of the exposure module.
Note
The Auto-Pak cloth will not remove large particles. Use a flashlight and fingers to
remove large particles.