User Manual

Operation and Maintenance
June 12, 2003 5E6155 3-21
7. Discard the 2 used filters.
Preventive Maintenance To maintain optimum performance from the laser imager, Kodak
recommends that preventive maintenance is completed by a Kodak
authorized Service Provider after every 20,000 film cycles.
The Status screen displays a “Preventive Maintenance” icon when the film
count reaches 20,000 sheets of film. You can also monitor the “Prints to
PM” line on the System Information screen.
Preventive Maintenance includes:
Installation of new charcoal filters
Image assembly cleaning
Film processor cleaning
Call Service to schedule preventive maintenance (see “Calling for Support”
on page 4-25).