Specification Sheet

LCD INDICATION
EL-USB-2-LCD
Humidity, Temperature and Dew Point Data Logger with LCD Display
Specications liable to change without prior warning Issue 7 07/2012 S.L Applies to EL-USB-2-LCD Page 3 of 5Specications liable to change without prior warning Issue 7 07/2012 S.L Applies to EL-USB-2-LCD Page 2 of 5
www.lascarelectronics.com
RH%
Red/
Green LED
°C (°F)
Red/
Green LED
Green double ash
The data logger is not currently logging, but is primed to start at a later date and time (delayed start).
Green single ash
The data logger is currently logging. No alarm on the channel.
Red single ash
The data logger is currently logging. Low alarm on the channel.
Red double ash
The data logger is currently logging. High alarm on the channel.
Green triple ash
The data logger is full and has stopped logging. No alarm on the channel.
Red triple ash
The data logger is full and has stopped logging. Alarm (high, low or both).
No LEDs ash
The data logger is stopped, the battery is empty or there is no battery tted.
Dual Red ash (every 60 seconds)
The data logger battery is running low as its voltage has dropped below 2.9V
10 seconds later
RH%
°C (°F)
RH%
°C (°F)
Delayed Start
Push to Start
Logging
Stopped
This is shown when the logger is set to
start at a specic data and time*
This is shown when the logger is setup
for “Push to start” logging
This is shown when the logger is running
in “LCD off” mode, and the button is
pressed. The display clears again after
three seconds
If the logger has not been set to log and
the button is pressed, three dashes are
displayed for three seconds
* If the logger is set to “LCD off” or “LCD on for 30 seconds” mode, then this will only be shown after the button
is pressed. Otherwise the display will remain blank.
LED FLASHING MODES
EL-USB-2-LCD features 2 bi-colour LEDs; one LED represents temperature measurement, the other represents RH. Each is clearly marked
on the logger. To save power, the status indication alternates between the two channels every 10 seconds. First you will see the status of
the temperature logging, and 10 seconds later you will see the status of the RH logging, and so on.
Display Logger Status Explanation