Troubleshooting guide

I have done extensive testing on bulbs and units and have found this to be an
increasing problem. The reason for this appears to be that most all UV Lamps are now
manufactured for electronic ballasts not for the old transformer ballast systems. I have
personally tested well over 100 bulbs from multiple shipments and suppliers to find that
most of them (although a few do) will not work in a ballast system and DO work in the
electronic system.
My conclusion
The age of these units, combined with the latest manufacturing methods used has
caused an incompatibility between the bulbs we have to sell and the equipment. Either
the equipment or ballast is “just worn out enough” to not work, or the bulbs just don’t
work with them.
I continue to search for a solution for this issue. For now, I have plans to provide an
upgrade kit for your unit that allows you to switch over to an electronic ballast and avoid
having to put your unit down. If all goes will, this will happen by September 2014. So, if
it’s after that date, check www.ecoquestparts.com for that upgrade kit.