User's Manual

Acclarix AX3 Series Diagnostic Ultrasound System User Manual Exam Data Management
- 96 -
9 Exam Data Management
9.1 Storing Images
The system supports storing static images and cine clips. What is displayed in Information area,
Image area and Image parameter area on the main screen will be stored.
Storing static images:
Pressing the hard key on the console will always capture what is on the image area of the
screen. This includes live, frozen, or Cine images. It also includes reports or other GUI screens and
review.
Storing a clip:
Pressing the hard key on the console will capture the moving images in scanning or cine
review status.
Live Store
Live store is to capture the moving images in image scanning status. The system continues scanning
while storing.
The store starts with the press of the hard key and continues for the configured length of the
clip or until the clip store is interrupted. The length of the clip can be configured by setting the
Duration for prospective on the Store/Print page (See section 11.1.3 for details).
Frozen Store
Frozen store is to store the cine clip between the start frame and end frame when in cine review
status.
Freeze the image and press the hard key to start storing. The default length of the cine clip
between start frame and end frame can be configured by setting the Duration for retrospective on
the Store/Print page(See section 11.1.3 for details). Or, you can manually change the start frame or
end frame to determine the clip length.