User's Manual

Acclarix AX3 Series Diagnostic Ultrasound System User Manual Troubleshooting and Maintenance
- 124 -
Troubleshooting for high system temperature:
1. Check whether system vents are clear and unobstructed. See section 3.2.1 for the location of
system vents. Too much dust on the vents or obstructed vents could cause high temperature to
the system.
2. Check whether the ambient temperature where the system is used is above 40. The system
must be used in the ambient temperature lower than 40.
3. Check whether the five fans at the system vents are working.
If all these troubles above are cleared, but the system still displays high temperature warning, please
contact the Service.
13.3 Cleaning the System
Use only the EDAN-approved substances and methods listed in this chapter to clean the system. The
warranty does not cover damage caused by using unapproved substances or methods.
Edan Instruments has validated the cleaning and disinfection instructions included in this User Manual.
It is the responsibility of the healthcare professional to ensure that the instructions are followed so as
to ensure adequate cleaning and disinfection.
General Points:
Keep your monitor, cables and accessories free of dust and dirt. To prevent the device from damage,
please follow the procedure:
Use only recommended cleaning substances and disinfectants listed in this manual. Others may
cause damage (not covered by warranty), reduce product lifetime or cause safety hazards.
Always dilute according to the manufacturer‟s instructions.
Unless otherwise specified, do not immerse any part of the equipment or any accessories in
2.
Strip-shape or snowflake-shape
disturbance occurs on the display
screen.
1. Inspect the power supply.
2. Check whether it is disturbed by the ignition action
of any other device.
3. Check the disturbance of electric or magnetic field
in the surrounding environment.
4. Check whether the plug and socket of power supply
and transducer are properly connected.
3.
Image is not displayed clearly on
the screen.
1. Adjust overall gain (Gain).
2. Adjust eight TGC slide controls.
3. Adjust focus (the number and the position).
4.
Image window is dark.
1. Adjust the brightness and slide the TGC controls on
the touch screen.
2. Check whether the transducer is connected
correctly.
5.
The button is unresponsive
1. Check the control panel to see whether the button
is blocked and press it several times to release it.
2. Clean the button