User's Manual

Acclarix AX3 Series Diagnostic Ultrasound System User Manual Troubleshooting and Maintenance
- 125 -
liquid.
Do not pour liquid onto the system.
Do not allow liquid to enter the case.
Never use abrasive material (such as steel wool or silver polish).
Inspect the monitor and reusable accessories after they are cleaned and disinfected.
WARNING
The console is not waterproof. Do not immerse or expose to extended moisture. Splash resistance
does not extend to transducer connectors. Keep connectors dry.
13.3.1. System Cleaning and Disinfection
The validated disinfecting agents for cleaning and disinfecting the system are:
Ethanol (75%)
Isopropanol (70%)
To clean/disinfect the system surface:
1. Switch off the system and unplug it.
2. Use a soft cloth dampened with disinfecting agent to gently wipe the entire exterior surface,
including the screen, of the equipment thoroughly until no visible contaminants remain.
3. Wipe off the disinfecting agent with a fresh cloth or towel, dampened with tap water after cleaning
until no visible cleaning agent remains.
4. Dry the system in a ventilated and cool place.
NOTE:
1. Make sure the system is free of gel and any other visible residue.
2. Use a soft dry cloth without chemicals for cleaning. The monitor surface is easily scratched.
To clean/disinfect the trackball:
1. Remove the front panel bezel.
2. Remove the trackball as shown in
Figure 13-1
.
3. Wipe the trackball, X and Y rollers, and the auxiliary idler wheel with a soft cloth dampened with
the disinfecting agent until no visible contaminants remain.
4. Wipe off the disinfecting agent with a fresh cloth or towel dampened with tap water until no visible
disinfecting agent remains.
5. Wipe off with a dry cloth to remove residual moisture.
6. Leave the trackball, X and Y rollers, and the auxiliary idler wheel to air dry.
7. Assemble the trackball and front panel bezel after the assembly parts completely dry.