User's Manual

Acclarix AX3 Series Diagnostic Ultrasound System User Manual EMC Information
- 140 -
Electromagnetic immunity
Guidance and manufacture’s declaration – electromagnetic immunity
The system is intended for use in the electromagnetic environment specified below. The customer or
the user of the system should assure that they are used in such an environment.
Immunity test
IEC/EN 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC/EN
61000-4-6
Radiated RF
IEC/EN
61000-4-3
3 V
rms
150 kHz to 80 MHz
6Vrms
c
in ISM bands
between
0.15 MHz and 80
MHz
3 V/m
80 MHz to 2.7 GHz
See table 1
3 V
rms
150 kHz to 80
MHz
6Vrms
c
in ISM
bands
between
0.15 MHz and
80 MHz
3 V/m
80 MHz to 2.7
GHz
Comply with
table 1
Portable and mobile RF communications
equipment should be used no closer to
any part of the system, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
Pd 2.1
150KHz to 80MHz
Pd 2.1
80 MHz to 800 MHz
Pd 3.2
800 MHz to 2.7 GHz
at RF wireless
communications equipment bands
(Portable RF communications
equipment (including peripherals such
as antenna cables and external
antennas) should be used no closer than
30 cm (12 inches) to any part of the
system, including cables specified by the
manufacturer).
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol: