User's Manual

Acclarix AX3 Series Diagnostic Ultrasound System User Manual Getting Started
- 27 -
3.3.4. Powering on/ off
To power on:
1. Connect the system to a hospital grade power supply, or use the battery as the power supply.
2. Press the Power on/off key on the top left of the control panel.
To power off:
1. Press the Power on/off key on the top left of control panel and the system displays a
confirmation dialog box.
2. Select “Shut Down” from the confirmation dialog box.
If the system is unresponsive, a long press of the Power on/off key will shut down the system directly.
NOTE:
1. Turn off and unplug the device after use.
2. Please unplug the AC adapter from the power socket and disconnect the battery prior to storage.
Sleep mode
The system will enter a sleep mode that maintains exam information while using minimal power.
There are three events that can invoke sleep mode:
Close the cover of the system without powering-off the system.
No user input for a configurable amount of time. Please see System Set-up to configure this
time.
Pressing the Sleep button on the confirmation dialog box when powering off.
Figure 3-10 Confirmation Dialog Box When Power off
There are two events that can exit sleep mode:
Open the cover of the system.
Press the Power on/off key.