User's Manual

SE-1010 PC ECG User Manual Hint Information
- 147 -
Sampling Device is in sleep mode, please
press “Power” to wake it up.
DX12 transmitter is in sleep mode.
Overload
The direct current offset voltage on an electrode
is too high.
Sorry, Can not Connect to the Database!
MSDE 2000 or SQL Server 2005 Express is not
started up.
Fail to create database! The system fails to create database.
Fail to open the treadmill (ergometer)
port! Please make sure the treadmill
(ergometer) has been connected to the
computer and the port setting is right
Fail to open the COM port which controls the
treadmill/ergometer. Reconnect the
treadmill/ergometer to the PC, and set the COM
port correctly.
Fail to open the BP monitor port! Please
make sure the BP monitor has been
connected to the computer and the Port
set is right!
Fail to open the COM port which controls the
blood pressure. Reconnect the BP monitor to the
PC, and set the COM port correctly.
The current HR has exceeded the target
HR!
Current heart rate value exceeds the target heart
rate value.
The diastolic BP has exceeded the normal
range!
Diastolic blood pressure exceeds the normal BP
range.
The systolic BP has exceeded the normal
range!
Systolic blood pressure exceeds the normal BP
range.
According to IEC/EN 60601-1-8, the characteristics of the visual alarm signals (hint information)
are listed in Table 10-2.
Table 10-2 Characteristics of Visual Alarm Signals (Hint Information)
Alarm Category Indicator Color Flashing Frequency Duty Cycle
LOW Yellow Constant (on) 100%