User's Manual

SE-1010 PC ECG User Manual EMC Information
- 161 -
Guidance and manufacture’s declaration - electromagnetic immunity -
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration - electromagnetic immunity
SE-1010 PC ECG is intended for use in the electromagnetic environment specified below. The
customer or the user of SE-1010 PC ECG should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
Not applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line to line
±2 kV line to groud
Not applicable Mains power quality
should be that of a typical
commercial or hospital
environment.
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
3A/m 3A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
Not applicable Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
SE-1010 PC ECG requires
continued operation during
power mains interruptions,
it is recommended that