User's Manual

SE-1010 PC ECG User Manual EMC Information
- 162 -
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
SE-1010 PC ECG be
powered from an
uninterruptible power
supply or a battery.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacture’s declaration - electromagnetic immunity -
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration - electromagnetic immunity
SE-1010 PC ECG is intended for use in the electromagnetic environment specified below. The
customer or the user of SE-1010 PC ECG should assure that it is used in such an environment.
Immunity
test
IEC 60601 test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3V
rms
3 V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of SE-1010
PC ECG, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
P
V
d
=
1
5.3
P
E
d
=
1
5.3
80 MHz to 800 MHz
P
E
d
=
1
7
800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the