User's Manual

SE-1515 PC ECG User Manual Technical Specifications
- 96 -
A1.5 Performance Specifications
HR Recognition
HR Range
30 bpm ~300 bpm
Accuracy
±1 bpm
ECG Sampling Box Performance
Leads
DE18
18 leads
DE15
16 leads
DP12/DX12
12 leads
Acquisition Mode
DE18
simultaneously 18 leads
DE15
simultaneously 16 leads
DP12/DX12
simultaneously 12 leads
Sample Frequency
DE18/DE15
1kHz (Analysis)
16kHz (Sampling)
DP12
1kHz (Analysis)
1kHz (Sampling)
DX12
0.5kHz (Analysis)
10kHz (Sampling)
A/D
DE18/DE15/DP12
24 bits
DX12
18bits
Resolution
DE18/DE15
0.1575uV/LSB
DP12/ DX12
2.52uV/LSB
Input Voltage Range
<±5mVp-p
Time Constant
≥3.2s