User's Manual

SE-1515 PC ECG User Manual EMC Information
- 99 -
Appendix 2 EMC Information
Electromagnetic Emissions
Guidance and manufacture's declaration - electromagnetic emission
SE-1515 PC ECG is intended for use in the electromagnetic environment specified below. The
customer or the user of SE-1515 PC ECG should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
SE-1515 PC ECG uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class A
SE-1515 PC ECG is suitable for use in all
establishments, other than domestic and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Electromagnetic Immunity
Guidance and manufacture's declaration - electromagnetic immunity
SE-1515 PC ECG is intended for use in the electromagnetic environment specified below. The
customer or the user of SE-1515 PC ECG should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.