User's Manual

SE-1515 PC ECG User Manual EMC Information
- 100 -
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
Not applicable
Not applicable
Surge
IEC 61000-4-5
1 kV line to line
2 kV line to ground
Not applicable
Not applicable
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
Not applicable
Not applicable
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Electromagnetic Immunity
Guidance and manufacture's declaration - electromagnetic immunity
SE-1515 PC ECG is intended for use in the electromagnetic environment specified below. The
customer or the user of SE-1515 PC ECG should assure that it is used in such an environment.