SE-1515 PC ECG User Manual Preparations Before Operation
- 34 -
3.4.3.2 Attaching the Chest Electrodes
Chest Electrode
Chest Electrode Connection:
1) Ensure that the electrodes are clean;
2) Clean the electrode area on the chest surface with 75% alcohol;
3) Daub the round area of 25mm in diameter on each electrode site with gel evenly;
4) Place a small amount of gel on the brim of the chest electrode's metal cup;
5) Place the electrode on the chest electrode site and squeeze the suction bulb. Unclench it and
the electrode is adsorbed on the chest;
6) Attach all chest electrodes in the same way.
NOTE: Long-time measurement with a strong negative pressure on the suction bulb may
cause reddening of the skin. When using the electrode on kids or patients with
delicate skin, squeeze the suction bulb lightly.
3.4.4 Attaching Disposable Electrodes
Disposable Electrode (clip style): Alligator Clip:
Disposable Electrode Connection (Clip Style)
1) Align all lead wires of the patient cable to avoid twisting, and connect the alligator clips to
the patient cable.
Suction Bulb
Connecting to a Lead Wire
Metal Cup