User's Manual

SE-1515 PC ECG User Manual Preparations Before Operation
- 35 -
2) Clean the electrode areas on the body surface with 75% alcohol.
3) Attach the disposable electrodes to the electrode positions on the body surface.
4) Clip the disposable electrodes with the alligator clips.
Snap/Banana Socket Adapters Disposable Electrode (Snap Style)
Disposable Electrode Connection (Snap Style)
1) Align all lead wires of the patient cable to avoid twisting, and cnnect Snap/Banana
Socket Adapters to connector of patient cable.
2) Clean the electrode areas on the body surface with 75% alcohol.
3) Attach the disposable electrodes to the electrode positions on the body surface.
4) Connect Snap/Banana Socket Adapters to the disposable electrodes.
WARNING
1. The disposable electrodes can only be used for one time.
2. The clip electrodes should be used together with the alligator clips.
3.5 Inspection Before Test
In order to avoid safety hazards and get good ECG records, the following inspection procedure is
recommended before operation.
1) Environment:
Make sure that there is no electromagnetic interference source around the equipment,
especially large medical electrical equipment such as electrosurgical equipment,
radiological equipment, magnetic resonance imaging equipment etc. Switch off these
devices when necessary.
Keep the examination room warm (above 18) to avoid muscle action voltages in
ECG signals caused by cold.