User's Manual

SE-1515 PC ECG User Manual ECG Sampling
- 43 -
Standard+NEHB
І, II, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6, ND,
NA, NI
Standard+XYZ
І, II, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6, X, Y,
Z
Children mode
І, II, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6, V3R,
V4R, V7
Under the 16-lead mode, you can choose Standard+Right.
Lead Sequence
Lead Group
Standard+Right
І, II, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6, V3R,
V4R, V5R, V7
Under the 18-lead mode, you can choose Standard+Right+Back.
Lead S quence
Lead Group
Standard+Right+Back
І, II, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5,
V6, V3R, V4R, V5R, V7, V8, V9
Lead electrode
Options provided are IEC and AHA.
You can configure it based on the patient cable used.
Display Order
Options provided are Simultaneously and Continuously.
If continuously is selected, the lead group will be displayed group by group.
If simultaneously is selected, all the leads will be displayed at the same time.
QRS Sound
Options provided are ON and OFF.
AC Filter
It is used to protect on the ECG signals from the interferences of the AC power.
Options provided are OFF, 50Hz, and 60Hz.
Select 50Hz/60Hz, AC50/AC60 will be printed on the report respectively.
DFT Filter
It is used to ensure that the ECG signals are on the same level during sampling.
The value configured for DFT filter will be printed on the report.
Auto
Diagnosis
Options provided are Display Standards, Display Normal ECG only, Display
All and OFF.
If Display Standards is selected, the system generates the automatic diagnosis
result after the sampling is complete.
If Display Normal ECG only is selected, the system only generates the
diagnosis result for the normal ECG after the sampling is complete.
If Display All is selected, the system generates automatic diagnosis results for