SE-1515 PC ECG User Manual ECG Sampling
- 45 -
The wave width can be set to 1, 2, 3, 4, or 5.
A larger value means that the waveforms printed are wider.
NOTE: This parameter affects the printing only, the waveforms
displayed on the screen remains the same.
Print Setting
Sequence can be set to sequential or synchronous.
If sequential is selected, the lead group will be sampled group by
group.
If synchronous is selected, all the leads will be sampled at the
same time.
Paper Orientation can be set to Landscape or Portrait.
Paper Size can be set to A4 or Letter.
Baseline Adjustment can be set to OFF, Auto, or Horizontal.
If it is set to Horizontal, the baseline for leads on the same level is
in the same line.
If it is set to Auto, the system automatically adjusts the baseline
for each lead group.
If it is set to OFF, the system adjusts the baseline for each lead
group to an average value.
If Auto Gain Control (AGC) is selected, the gain will be adjusted
automatically, and the Baseline Adjustment is set to Horizontal
automatically.
If the printing color is set to color, the background grid will be printed
in color.
NOTE: If the printing color is set to color, but a black-and-white
printer is used, the report printed will be illegible.
If Print after diagnosis is selected, the system automatically prints the
report after the diagnosis is complete.
4.4.1.6 Function/Algorithm Setup
Click on Setting on the resting ECG sampling screen, and the System Setting window will be
displayed. In the System Setting window, click Function/Algorithm Setting to enter the
Function/Algorithm Setting screen.