User's Manual

SE-1515 PC ECG User Manual ECG Sampling
- 49 -
(X<10), the system will print the waveforms in the last Xs.
Freeze
F6
During the presampling stage and test, Freeze is displayed. Click Freeze and
the frozen window will be displayed.
Edit/Event
F7
During presampling, you can click Edit to open the New Patient window
and edit patient information.
During the test, Event is displayed. You can click it to mark an event.
BP
F8
Click BP in the auto measurement mode, the system will activate the BP
monitor and start measuring.
Click BP in the manual measurement mode, you can manually input BP in the
displayed window.
Start/Stop Tmill
F9
Only available in the Exercise and Recovery stage.
During the exercise test, you can click Stop Tmill to cool down the treadmill
and stop the exercise test temporarily, and click Start Tmill to continue the
exercise test.
Keep
Click Keep during the exercise test, and the system will maintain the current
speed and slope before you click Keep again.
NOTE: In the Keep status, the test cannot move into the next stage
automatically.
Next
Click Next during the exercise test and the system will move into a new stage.
NOTE: When the Pretest has lasted for less than 15s, you cannot click
Next to move from the Pretest to Exercise.
Speed Up/Speed
Down
You can click Speed Up/Speed Down in the Exercise stage to raise/reduce
the speed of the treadmill.
You can click Speed Up/Speed Down in the Exercise stage to raise/reduce
the output of the ergometer.
Slope Up/Slope
Down
You can click Slope Up/Slope Down in the Exercise stage to raise/reduce the
slope of the treadmill.
Exercise
Amount
Up/Down
You can click Exercise Amount Up/Down in the Exercise stage to
add/reduce 5W exercise amount at one time.