User's Manual

SE-1515 PC ECG User Manual ECG Sampling
- 52 -
4.4.3.6 Device Setup
Click on Settings on the exercise ECG sampling screen, and the System Setting window will be
displayed. In the System Setting window, click Device to enter the Device window.
Parameter
Description
Device Type
Options provided are Treadmill or Ergometer.
Device Model
You can configure the device model based on the device used.
If RUNNER RUN-2011/T treadmill is to be used, set the Device Model
to Trackmaster Treadmill.
Protocol
You can configure the protocol based on the device used or customize a
protocol.
You can click Edit to edit the protocol. For details, see section 4.5.3.7.
You can click Restore to factory defaults and the system will to reset
the protocols to the default value.
BP monitor
You can set BP monitor to the port that can be used on the computer.
BP Sampling Mode
Options provided are:
Once per stage
The system starts measuring the blood pressure immediately after
entering a new stage each time. However, in the Pretest stage, the
system only starts the blood pressure measurement in the first 10s of
the first stage.
Once every three/five/seven minutes
After the first time the blood pressure is measured in the pretest
stage, that is, the first 10s of the first stage, the system starts the
measurement every 3/5/7 minutes.
Protocol-Based Control
The blood pressure measurement starts as configured in the
Protocol Edit screen.
Manual Input
You have to input the blood pressure data manually.
Triggering mode
It can be set to Square wave or QRS.
Unit
The unit of Speed can be set to mph or km/h.
The unit of Slope can be set to % or degree.
ECG Key
It can be set to Pretest, Print, or Forbidden.