User's Manual

PADECG User Manual Operation instructions
- 23 -
Long pressing the waveform on the ECG Analysis screen can amplify the ECG waveform.
Press on the Amplified Waveform screen to measure the ECG waveform. Press
this button again to cancel measuring.
3. Modifying measure information
Long press one parameter in the top left corner of the ECG Analysis screen, and the Modify
measure information window pops up. Make the related settings, and then click on the OK
button to save the modifications.
The common parameters are displayed as follows.
Designation
Description
HR (bpm)
Heart Rate
P (ms)
P-wave duration of the current lead
PR (ms)
P-R interval of the current lead
QRS (ms)
QRS complex duration of the current lead
QT/QTc (ms)
Q-T interval of the current lead/Normalized QT interval
P/QRS/T (° )
Dominant direction of the average integrated ECG vectors
RV5/SV1 (mV)
The amplitude of R wave of V5 lead/the amplitude of S wave of
V1 lead
RV5+SV1 (mV)
The amplitude of R wave of V5 lead plus the amplitude of S wave
of V1 lead
RV6/SV2 (mV)
The amplitude of R wave of V6 lead/the amplitude of S wave of
V2 lead
4. Modifying diagnosis results
Long press one diagnosis result in the top right corner of the ECG Analysis screen, and the
Auto Diagnosis window pops up.
Select one diagnosis result from the Diagnosis List or input diagnosis information directly
in the textbox, and click on the OK button to save the modifications.
5. Modifying Diagnosis List
Long press one diagnosis result in the Diagnosis List, and you can add, delete, or modify
the diagnosis result.
NOTE: If Diagnosis List is modified, you should click on the Save button in the
Auto Diagnosis window to save the related modifications.