User's Manual

PADECG User Manual EMC Information
- 41 -
Appendix 2 EMC Information
Electromagnetic Emissions
Guidance and manufacture’s declaration - electromagnetic emission
The PADECG is intended for use in the electromagnetic environment specified below. The
customer or the user of the PADECG should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The PADECG uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emission
CISPR 11
Class A
The PADECG is suitable for use in all
establishments, other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Electromagnetic Immunity
Guidance and manufacture’s declaration - electromagnetic immunity
The PADECG is intended for use in the electromagnetic environment specified below. The
customer or the user of the PADECG should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
8 kV contact
15 kV air
8 kV contact
(only for
PADECG)
15 kV air (only
for PADECG)
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.