Use Instructions

Vital Signs Monitor User Manual Product Specification
- 96 -
A Product Specification
NOTE:
The performance of the equipment with mark is determined to be essential
performance.
A.1 Classification
Anti-electroshock type
Class I equipment and internal powered equipment
Anti-electroshock degree
SpO
2
, NIBP: CF
TAT TEMP, F3000 TEMP: BF
Ingress Protection
Monitor: IP44 (protected against splashing water and solid
foreign objects 1.0 mm diameter)
CS-04 Charger Stand: IPX1 (Protected against vertically
falling water drops)
CS-05 Extended Stand: IPX1 (TAT TEMP, F3000 TEMP
have no ingress protection)
Disinfection/sterilization method
Refer to Chapter Care and Cleaning for details.
Working system
Continuous operation equipment
Compliant with Standards
IEC 60601-1: 2005+A1: 2012; IEC 60601-1-2: 2014;
EN 60601-1: 2006+A1: 2013; EN 60601-1-2: 2015;
EN 80601-2-30: 2010+A1: 2015; EN 60601-2-49: 2015;
EN ISO 80601-2-56: 2017+A1:2018;
EN ISO 80601-2-61: 2017
A.2 Physical Specifications
A.2.1 Size and Weight
Size (monitor)
(77±1) mm (W) × (150±1) mm (H) × (28±1) mm (D)
Weight (monitor)
< 300 g (standard configuration, without battery and
accessories)
< 350 g (standard configuration, with rechargeable battery,
without accessories)
< 370 g (standard configuration, with AA battery, without
accessories)