Use Instructions

Vital Signs Monitor User Manual Product Specification
- 100 -
A.3 NIBP
Technique
Oscillometry
Mode
Manual, Continuous, Average
Continuous
5 min, interval is 5 s
Measuring Parameter
SYS, DIA, MAP, PR
Pressure Unit
kPa, mmHg
Average measurement
Interval (unit: minutes)
1/2/3/4/5
Times
3/5
Measuring Range
Adult Mode
SYS: 25 mmHg to 290 mmHg
DIA: 10 mmHg to 250 mmHg
MAP: 15 mmHg to 260 mmHg
Pediatric Mode
SYS: 25 mmHg to 240 mmHg
DIA: 10 mmHg to 200 mmHg
MAP: 15 mmHg to 215 mmHg
Neonatal Mode
SYS: 25 mmHg to 140 mmHg
DIA: 10 mmHg to 115 mmHg
MAP: 15 mmHg to 125 mmHg
Cuff Pressure Measuring
Range
0 mmHg to 300 mmHg
Pressure Resolution
1 mmHg
Maximum Mean Error
±5 mmHg
Maximum Standard Deviation
8 mmHg
Maximum Measuring Period
Adult/Pediatric
120 s
Neonate
90 s
Typical Measuring Period
30 s to 45 s (depend on HR/motion disturbance)
Dual Independent Channel Overpressure Protection
Adult
(297±3) mmHg
Pediatric
(245±3 ) mmHg
Neonatal
(147±3) mmHg