Vital Signs Monitor User Manual Intended Use and Safety Guidance
- 4 -
WARNING
32 SHOCK HAZARD - Don't connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlet supplying the system.
33 SHOCK HAZARD - Don't connect electrical equipment, which has been supplied as a
part of the system, directly to the wall outlet when the non-medical equipment is
intended to be supplied by a multiple portable socket-outlet with an isolation
transformer.
34 Operation of the equipment exceeding specified physiological signal or the
operational specification may cause inaccurate results.
35 To protect the monitor from damage during defibrillation, for accurate measurement
information and to protect against noise and other interference, use only accessories
specified by EDAN.
36 No modification of this equipment is allowed without authorization of the manufacturer.
If this equipment is modified, appropriate inspection and testing must be conducted to
ensure continued safe operation.
37 Clinical decision making based on the output of the device is left to the discretion of
the provider.
38 The monitor is equipped with Wi-Fi to receive RF electromagnetic energy. Therefore,
any other equipment complying with CISPR radiation requirements may also interfere
with the wireless communication and make it interrupted.
39 Wireless LAN equipment contains an intentional RF radiator that has the potential of
interfering with other medical equipment, including patient implanted devices. Be sure
to perform the electromagnetic compatibility test before installation and any time new
medical equipment is added to the Wireless LAN coverage area.
40 If the protective grounding (protective earth) system is doubtful, the monitor must be
supplied by internal power only.
41 Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the monitor, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
42 Make sure networking function is used in a secure network environment.