Use Instructions

Vital Signs Monitor User Manual EMC Information
- 106 -
B EMC Information
- Guidance and Manufacture’s Declaration
B.1 Electromagnetic Emissions
Guidance and manufacture’s declaration – electromagnetic emission
iM3s is intended for use in the electromagnetic environment specified below. The customer or
the user of iM3s should assure that they are used in such an environment.
Emission test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
iM3s uses RF energy only for their internal
function. Therefore, their RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
RF emission
CISPR 11
Class B
(Class A: when
work with CS-05
Extended Stand)
iM3s is suitable for use in all establishments,
other than domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
Complies
NOTE:
(When work with CS-05 Extended Stand) The EMISSIONS characteristics of iM3s make
it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B is normally required) iM3s might not
offer adequate protection to radio-frequency communication services. The user might
need to take mitigation measures, such as relocating or re-orienting the equipment.
B.2 Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
iM3s is intended for use in the electromagnetic environment specified below. The customer or
the user of iM3s should assure that they are used in such an environment.
Immunity test
IEC/EN 60601 test level
Compliance level
Electromagnetic
environment -
guidance